BT-700 Operator Manual 25
P/N : 700-ENG-OPM-EUR-R02
Bistos Co.,Ltd.
2018,03
Section
8
Troubleshooting and
Maintenance
8.1
Sensors
To test an Sensors:
1.
Properly connect the Sensors to the right of the monitor.
2. Turn on the monitor.
3. Adjust the speaker volume to an audible level.
4. The Pulse Search indicator is lit for more than 10 seconds while
the sensor is connected to the patient.
5. The Pulse Search indicator lights after successful measurements
have been made.
8.2
Battery Disposal and Handling
Be cautious when disposing of internal Ni-MH battery.
Adhere to all applicable laws regarding recycling. Avoid storing battery
above 60°C (140°F). If clothing or skin comes in contact with material from
inside the battery, immediately wash with plenty of clean water.
When you replace internal Ni-MH, contact to our technical support office.
8.3
Maintenance
BT-700 requires periodic calibration or adjustment. The recommended
interval for performing hipot and leakage testing is once per year.
If you need a Calibration or a adjustment, contact to our technical support
office.
8.4
Troubleshooting
(1)
If not Power on, change the power adaptor. And try again turn on the
power button.
(2)
If not displayed SpO2 signal, change the SpO2 Probe.
(3)
If not displayed Temperature value, change the temperature sensor.