Section
7
Cleaning and Disinfection
This chapter contains instructions for the care and cleaning of the BT-700 unit
and its accessories.
The BT-700 requires proper care and preventive maintenance. This ensures
consistent operation and maintains the high level of performance necessary in
monitoring procedures.
7.1
Monitor
Keep the external surface clean and free of dust, dirt, and residual liquids.
Clean with a damp cloth using mild soap and water or hospital approved
nonabrasive disinfectants.
7.2
Sensors
Cleaning and Disinfecting the SpO2 sensor and the temperature
sensor.
To avoid damage to the sensors, clean and disinfect only according to the
following instructions.
1. Wipe the device with a sterile wipe soaked in enzymatic detergent safe for
use with metal instruments. Wipe the exterior of the device three times.
Prepare the detergent according to the manufacturer’s sensor
recommendations.
2. Scrub sensors with enzymatic detergent using soft bristled brush for five (5)
minutes. DO NOT IMMERSE SENSORS.
WARNING
: Unplug the monitor from the AC power source and detach all
accessories before cleaning. Do not immerse the unit in water or allow liquids
to enter the case.
CAUTION
: Take extra care when cleaning the display surfaces, which are
sensitive to rough handling. Rub the lens that covers them with a soft, dry
cloth.
CAUTION
: Do not autoclave. Do not gas sterilize.
CAUTION
: Do not immerse in liquid. When using solutions, use sterile
wipes to avoid pouring fluids directly on the transducer.