Bistos Co.,Ltd.
2018,03
3. Wipe sensors three (3) times with sterile water to remove soap residue.
4. Wipe the sensors with a sterile wipe soaked in Cidex™. Wipe all exterior
surfaces of the transducer three (3) times.
5. Wipe the sensors three (3) times with sterile water to remove Cidex residue.
6. Dry the device thoroughly with a sterile soft towel or gauze surgical sponge.
7. Wrap the dry device in a fresh sterile soft towel or transparent sterile wrap
for storage until next use.
WARNING
: Before use, carefully read the sensor directions for use,
including all warnings, cautions, and instructions.
WARNING
: Do not use a damaged sensor or pulse oximetry cable. Do not
use a sensor with exposure optical components.
WARNING
: Use only one pulse oximetry cable to increase the length of the
sensor. Use of more than one pulse oximetry cable may have an adverse effect
on performance. Do not attach any cable that is intended for computer use to
the sensor port.
WARNING
: Pulse oximetry readings and pulse signal can be affected by
certain ambient environmental conditions, sensor application errors, and certain
patient conditions.
WARNING
: Tissue damage can be caused by incorrect application or duration
of use of the SpO2 sensor. Inspect the sensor site as directed in the sensor
directions for use.
WARNING
: Do not immerse or wet the sensor.