Troubleshooting 39
P/N 128097 Rev. 00
TurboVap 96 User’s Manual
Biotage AB
Workstation Not Operational
To diagnose problems with the unit, locate the description of the
condition in the table below.
Elapsed Time displays
“
”.
Gas Time is set to
‘infinite’ mode.
This is the normal display after
99 minutes.
LED out on front panel.
Faulty LED
Contact Biotage (see
page 3
).
Value or Units display
does not work.
Faulty display chip
Contact Biotage (see
page 3
).
Condition
Probable Cause
Solution
Condition
Probable Cause
Solution
Unit is inoperative.
Power is off, or the
power cord is loose.
Check that the power switch is
on and that the power cord is
seated.
Faulty fuse(s).
Replace the fuse(s).
Faulty control board.
Contact Biotage (see
page 3
).
Flow Rate is not
operational or not
correct.
Bad sensor PCB.
Contact Biotage (see
page 3
).