Hardware
18
P/N 128097 Rev. 00
TurboVap 96 User’s Manual
Biotage AB
Cycle Readout
- when lit, all of the parameters, one at a time, for
any position that is running are displayed. If neither position is
running, the display shows the set values for both positions. The
plate temperature display shows the present temperature when
evaporation is in process for that position; otherwise, it shows the
set value.
Select Button
Advances the display indicator to the next option. If you press the
Select button after turning on the workstation, the TurboVap enters
diagnostics mode, displaying the current firmware revision.
Set Button
This button sets currently selected setting (Gas Time or Plate
Temperature) to the displayed value. The Set button only sets the
Gas Time and Plate Temperature. The Set button has no effect on
the other settings.
Up and Down Arrow Buttons
Increase or decrease the value in the display. Used to set the
values for Gas Time and Plate Temperature.
Start
Stop Buttons
Start or stop evaporation. During evaporation, the Position light is
green.
Over Temp Light
Turns on (red) when the present temperature of one of the position
is 10
C greater than the set plate temperature. When the
temperature drops below this value, the light turns off. The light may
indicate a heater runaway caused by a defective relay or
temperature sensor.
Low Flow Light
Indicates a low gas flow rate, possibly because the gas is running
out, or the flow regulator was not adjusted before starting the
evaporation. If the flow rate drops below 5 SCFH when the
evaporation is in process, the Low Flow LED lights and the alarm
beeps continuously.