Installation
24
P/N 128097 Rev. 00
TurboVap 96 User’s Manual
Biotage AB
Connect the Power Cord
1
Plug the power cord securely into the receptacle on the back of
the unit. Make sure that the plug is fully inserted.
2
Connect the 3-pronged end to an appropriate power source.
WARNING
To avoid the risk of fire or electrical shock, plug the power cord into
a properly grounded outlet.
Turn On the Power
To turn on the TurboVap 96, turn ON the power switch located on
the lower left side of the unit.
When the power is turned on, the internal air temperature will heat
up to the previously set plate temperature if it was set to below
37°C; or it will heat up to 37°C if it is set above this value.
Power-Up Diagnostics
When the power is turned on, both the Value and Units displays
show 88, and all LEDs are lit for 1.5 seconds. Confirm that all lights
are working properly. (If you press the
Select
button while in this
state, the workstation enters diagnostics mode, displaying the
current firmware revision.) After 1.5 seconds, initialization
continues.
Also during this time, raising and lowering the cover sounds the
beeper to confirm the cover switch is functioning.
After initialization, the workstation lights the Position 1 – Gas Time
light and displays the Gas Time previously set for Position 1.