Installation
21
P/N 128097 Rev. 00
TurboVap 96 User’s Manual
Biotage AB
WARNING
To avoid injury to yourself or damage to the instrument, DO NOT
exceed 80 psi (5.52 bars) maximum supply pressure.
Exhaust Duct
The exhaust duct requires a 2-inch (5.1-cm) diameter venting hose
(supplied). The exhaust duct
MUST
be routed to a suitable
ventilation system vented
outside
the laboratory.
WARNING
Exhaust gases may be hazardous. Consult the Material Safety Data
Sheets (MSDS) for all of the solvents used.
Power
100/120 VAC Model
220/240 VAC Model
Input Voltage
100/120 VAC +/- 10%
220/240 VAC +/- 10%
Line
Frequency
50 - 60 Hz
50 - 60 Hz
Fuse
(1) 6A, F250V
(P/N 38904)
(2) 3.15A, F250V
(P/N 38902)
Ground
The 3-prong power cord plug supplies the
ground.