Prepare the System
Figure 8.
The Solvent Setup window.
Prime the Solvent Inlets
Before you start a purification on your system, you might need
to prime the solvent inlets to:
»
Remove any air bubbles from the pump and the
solvent inlets by flushing them with solvents.
»
Empty the solvent inlets of solvents used in the previous
purification and fill them with new solvents.
Note:
All four solvent inlets must be primed with solvent to
achieve the specified pump performance.
Note:
We recommend to have all solvent reservoirs on the same
physical height to improve accuracy in the solvent mixing.
1. In the software, select
Solvent Setup
from one of the solvent
drop-down lists in the gradient view. The
Solvent Setup
window opens; see Figure 8.
2. Prime the solvent inlets by pressing the corresponding
Prime
buttons. Note that 27 mL of the assigned solvent is
used for each prime.