Biotage
®
Selekt User Manual | © Biotage 2018
System Overview
Solvent Supply
The system is equipped with four solvent inlets found on the
right side of the system (S1-S4).
Optional Secondary Solvent Containment
A maximum of four 5-liter reservoirs can be placed on the
optional secondary solvent containment; see Figure 4.
Larger reservoirs than 5 liters must be placed elsewhere.
Note:
To be compliant with the secondary containment
regulations, do not use reservoirs larger than 4 liters.
Figure 4.
The optional secondary solvent containment.
Internal UV Detector
The system includes either a UV detector with the range 200 to
400 nm or a UV-VIS detector with the range 198 to 810 nm.
Collection and Fractionation Methods
Available detection signals are:
»
λ
-All:
The system uses average absorbance
within a user-defined wavelength range for
collection and fractionation. Possible collection
and fractionation parameters are Collect All,
Threshold, and Valley Fractionation. Only available
on systems with a Spektra software license.
»
UV1 and UV2:
The system uses one or two wavelength
signals for collection and fractionation. Possible
collection and fractionation parameters are Collect
All, Threshold, and Valley Fractionation.
All collection methods fractionate on volume, which is when
a test tube or bottle is full based on the specified vessel volume
for the rack type or a user-defined maximum fraction volume
defined for the run.
Signals that are not used for collection and fractionation can be
used for monitoring.
Audible Alarm
If the audible alarm is enabled in the system settings, an
audible warning will sound whenever user interaction is
required. Examples of situations are: 1) rack change is needed,
2) high pressure, 3) misalignment of the collection arm, and
4) pump failure.
Lighting
To provide better visibility for the user, there is a light
strip underneath the touch screen and a mirror behind the
collection racks.