Maintenance
8
Replacing the Detector Firmware
10
Select the firmware hex file and click
Open
.
The download will begin straightaway and the progress will be
displayed, as shown.
11
On successful completion of the file transfer, the software will
display a
Flash memory updated ok
message.
12
Power cycle the detector and check that the firmware version
displayed on the detector’s front panel at boot-up is correct.