50
Biotage
®
ELSD-A120 User Manual
6
Troubleshooting and Diagnostics
General Problems
Detector Fails to Zero
No Power
No Response (Completely Flat Baseline)
Probable cause
Suggested actions
1
Offset too high or output unstable
due to impurity in the mobile
phase
•
Stop the pump flow and switch off
the detector. Restart the detector
and auto-zero without liquid
flowing
•
Please contact your Biotage
service representative
•
Optical section contaminated and
requires cleaning
Probable cause
Suggested actions
1
Mains lead not connected
Attach the mains lead to the socket
and the inlet on the rear of the
detector
2
Fuse failure
Replace fuse
3
Power supply failure
Please contact your Biotage service
representative
Probable cause
Suggested actions
1
Data acquisition leads not
connected
Ensure that the connectors to the
Isolera system are sound
2
Light source inactive
•
Check that LED intensity is 100 %
•
Check that the LED is functioning
correctly, by stopping solvent
flow, cycling the power, and then
reading the offset value in RUN
mode
3
Output below 0 mV
Stop the pump flow and auto-zero
without liquid flowing
Summary of Contents for ELSD-A120
Page 1: ...Biotage ELSD A120 User Manual...
Page 12: ...12 Biotage ELSD A120 User Manual 1 Introduction to the Detector Detector Overview...
Page 44: ...44 Biotage ELSD A120 User Manual 5 Optimizing Performance Solvent Recommendations...
Page 60: ...60 Biotage ELSD A120 User Manual 7 Error Information Error Messages...
Page 74: ...74 Biotage ELSD A120 User Manual 8 Maintenance Replacing the Detector Firmware...
Page 85: ......
Page 86: ......