Error Information
Error Messages
Rear Vapor Sensor Failed
Error ID: 13
The vapor sensor located on the rear panel of the detector has failed.
Vapor Detected
Error ID: 14
Solvent vapor threshold exceeded inside the detector.
Leak Detected
Error ID: 15
A leak was detected inside the detector.
Probable cause
Suggested actions
1
Vapor sensor not connected to the
main board
Please contact your Biotage service
representative
2
Defective vapor sensor
Please contact your Biotage service
representative
Probable cause
Suggested actions
1
External vapor being drawn into
the detector
Remove any source of solvent vapors
close to the detector
2
Solvent leak inside the detector
Please contact your Biotage service
representative
3
Exhaust tube not fitted
Fit the black exhaust tube
Probable cause
Suggested actions
1
Loose nebulizer fittings
Please contact your Biotage service
representative
2
Blocked nebulizer causing leak at
capillary fittings
Please contact your Biotage service
representative