37
EMC TABLES
Immunity Test
CEI 60601
Test level
Compliance
level
Guidelines on the electromagnetic environment
Electrostatic Discharge
(DES)
CEI 61000-4-2
Power Supply
Frequency (50/60 Hz)
Magnetic Field
CEI 61000-4-8
± 6 kV
contact
± 8 kV air
3 A/m
± 6 kV
contact
± 8 kV air
3 A/m
The floors must be made of wood, cement or ceramic tiles. If the floors are
covered in synthetic material, relative humidity must be below 30%.
The magnetic field of the supply frequency must be at a level appropriate
to the place of use.
EMC TABLES
Emissions test
Compliance
Electromagnetic environment guidelines
RF Emissions
CISPR 11
RF Emissions
CISPR 11
Harmonic Emissions
CEI 61000-3-2
Voltage variations /
Flicker emissions
CEI 61000-3-3
Group 1
Class B
Not
applicable
Not
applicable
The 0081N wrist blood pressure monitor uses RF energy only for its internal operation.
Therefore, its RF emissions are very low and are not likely to cause interference in electronic
equipment nearby.
The 0081N wrist blood pressure monitor may be used in all establishments, including
dwellings and establishments directly connected to the public low-voltage power supply
system that powers buildings intended for homes.
Manufacturer’s declaration and directive – electromagnetic emissions
The 0081N wrist blood pressure monitor is intended for use in the electromagnetic environment
specified below. The client or user of the 0081N wrist blood pressure monitor must be sure to use it
in such an environment.
Manufacturer’s declaration and directive – electromagnetic immunity
The 0081N wrist blood pressure monitor is intended for use in the electromagnetic environment
specified below. The client or user of the 0081N wrist blood pressure monitor must be sure to use it
in such an environment.