Water Leaking.

The unit is tilt.

※ If the above solutions do not solve the problems, please contact BioSure distributor.

8

.TROUBLESHOOTING

Power is not connected.

Device doesn't work.

Use the power adapter attached in the product package.

Make sure the terminal of the power adapter is inserted 

into the power jack of the device.

If the main device is placed at an angle, the 

humidification water may leak out.

The REPLACE FILTER indicator 

lights up.

The life of DI water filter and 

HEPA filter have expired.

Please follow the instructions in section 7-1 to replace the 

DI water filter and HEPA filter.

The life of the EOG has expired. Please follow the instructions in section 7-3 to replace 

the EOG.

Insufficient of humidification 

water.

The WATER LOW indicator 

lights up.

The REPLACE CELL indicator 

lights up.

Control button(s) doesn't work.

Accurately touch the center of the operation button.

Did not touch the center of the 

operation button.

Replenish pure water or distilled water in accordance with 

the instructions in Section 5-5. If the indicator warning still

 

cannot be removed, please contact your service agent.

Insufficient pure water for the 

EOG.

The REFILL TANK indicator 

lights up.

The system can work normally. If you want to turn off the 

light, please power it up again.

When replenishing the pure 

water for the EOG, the user 

stops replenishing before the 

tank is filled.

The REFILL TANK indicator 

blinks.

The design of this product has fully considered safety, 

but some people may still feel a strong smell.

This is the smell of ozone.

I smell a strange smell at the 

ozone outlet.

Please contact your service agent.

The EOG fails. The unit stops 

running.

The SERVICE indicator lights up.

Replenish pure water to the humidification water tank 

according to the instructions in section 5-6.

Water quality: TDS <1ppm or EC <2μS/cm

15

Use the following guide to self-check your problems before requesting repairs.If you come into a conclusion that problem has 

occurred, do NOT attempt to disassemble and repair the product yourself, but contact BioSure or your dealer for support.

Sometimes suspected problem is remedied by disconnecting the power plug and then re-connecting it. 

Symptom

Possible Cause

Recommended Solution

Summary of Contents for PROFESSIONAL EOS7190-P

Page 1: ...es 5 Preparation Before Use 6 How to Use 7 Maintenance 8 Troubleshooting 9 Speci cations Warranty Description P2 P3 P5 P6 P7 P8 P8 P9 P10 P11 P12 P14 P15 P1 P16 Table of Contents Space Sanitizer Syste...

Page 2: ...EOS7190 P V2 1 EN...

Page 3: ...rying to repair or install 3 The labor costs derived from the normal maintenance calibration or general adjustment speci ed in this manual 4 Maintenance or damage caused by alteration abuse re ood or...

Page 4: ...cause personal injury 2 or property damage 3 if not avoided CAUTION indicates a dangerous situation that may cause personal injury 2 or property damage 3 if not avoided Do not use this product in plac...

Page 5: ...product The input power of the power adapter of this product is AC100 240V Do not use loose sockets Otherwise it may cause re electric shock or short circuit Do not use a damaged power cord or power...

Page 6: ...e do not use electrical electronic contact cleaners in addition to damaging to the appearance these chemicals may also melt the parts Do not rinse the main device with water If the main unit is dirty...

Page 7: ...REFILLING SMART SENSOR 12 9 11 10 1 3 4 5 6 7 2 8 FAN SPEED HUMIDIFIER GENERATOR TIMER LEVEL Mode Humidi cation Start Stop Timing Air Volume Water Replenishment Captions 4 1 Fault indicator 2 Standby...

Page 8: ...g to the Food and Drug Administration FDA According the Occupational Safety and Health Administration OSHA the concentration standard for continuous exposure to ozone is limited to exposure for 8 hour...

Page 9: ...sing please hang the product on the wall or place it on a rm and at table or shelf When ozone is generated the blue operation indicator on the top of the unit will be lit This light cannot be turned o...

Page 10: ...Water shut o valve Plug 6 Press the GENERATOR button to start the water pump The device will beep to start replenishing pure water At this time the REFILLING indicator will light up 7 After the water...

Page 11: ...he STANDBY indicator lights up and the device stops working Stop Operation After pressing the TIMER button to start running the system will rst run with the default value of low air volume for 2 hours...

Page 12: ...tering by mistake SMART SENSOR LEVEL FAN SPEED Beep x3 TIMER Beep Press again TIMER button Turn o the SMART SENSOR Restart the system 4 If you choose the Boost mode for safety reasons the device will...

Page 13: ...LACE FILTER indicator lights up stop the operation and replace the DI water lter and HEPA lter at the same time Click It must be replaced regularly to maintain the performance of the equipment When th...

Page 14: ...idifying water will be drained immediately after removing the silicone cover Discharge the humidi cation water and the pure water of the EOG Drain humidi cation water Drain pure water 2 3 1 Inside of...

Page 15: ...down the two tabs at the bottom and then remove the bottom cover 2 Separate the terminals of EOG and the main unit 5 Connect the terminals Inside the bottom cover Connection terminal Mark EOG 3 Rotat...

Page 16: ...ons in Section 5 5 If the indicator warning still cannot be removed please contact your service agent Insu cient pure water for the EOG The REFILL TANK indicator lights up The system can work normally...

Page 17: ...ns for use care or maintenance Removal alteration or defacing of the BES a xed serial number and other labels Mechanical damage Use of uids other than clean potable water This warranty is conditional...

Page 18: ...Memo...

Page 19: ......

Page 20: ...www besgroups com info besgroups com BES Group Biotek Environmental Science Ltd...

Reviews: