(5) Enter the patient s personal information including I.D. height, and gender
using the keypad buttons. After finishing the data entry, press the Enter
button and see the information windows prompting the patient to get ready for
a test.
(6) Follow the test instructions displayed on the information windows. InBody720
checks your posture on the equipment continuously. If the patient is settled on
the stand, and his/her posture is right, the testing commences on its own.
Once the test is underway, the patient should keep the same posture until the
end of the test.
If the patient doesn t take upon the right position, hold the hand electrode or step on the foot electrode
properly or if the patient s palms or soles are dry or have too much dead and hard skin, the testing
process may not initiate on its own. In this case, wipe up the patient s palm or feet with the electrolyte
tissues that come with the InBody720 and put the patient back on the testing stand.
If a data entered is out of the permissible data range, the following message will pop up
on the monitor. Go back to the test field where you were and re-enter data. Refer to the
5. Personal Profile of this chapter for the permissible range of each data.
If the patient s palms or soles are dry, the following message will be shown on the
information windows and the test stops. In this case, wipe up the patient s palms or
soles with the electrolyte tissues that come with the InBody720 and re-initiate the test
process.
Summary of Contents for InBody 720
Page 1: ...User s Manual...
Page 44: ...1 Setup Menu 2 Quick Setup 3 DATABASE 4 Modification Example Chapter 3 Setup Establishment...
Page 59: ...1 Consumables 2 Basic Equipment 3 Options Chapter 5 Consumables...
Page 71: ......
Page 72: ...InBody the product of great technology Experience its speciality...
Page 76: ...www biospace co kr...