If the patient s feet or hands are too dry, or has dead, hard skin built up, InBody720 may prompt the user
to re-test the patient. In this case, wet the palms and soles with electrolyte tissue that comes with the
InBody720 and re-test the patient.
When wetting hands and feet, keep the moisture on hands and feet from dripping to the electrodes.
Excessive moisture may cause an erosion of electrode, which in turn may result in breakdown of the
equipment. The electrolyte tissues provided by the Biospace are specifically manufactured for InBody and
thus is completely different from the generic wet tissues in the market. It is strongly recommended to use
the electrolyte tissues specifically designed for InBody720.
C. Body Posture
During testing, the LCD monitors display information on body composition,
allowing the patient to read the analysis from the stand. When the patient steps
down from the stand, the InBody720 goes back to the Initial screen.
(1) Do not leave the arms by your side. Form an angle of 15 degree between the
arms and your side.
(2) Stand comfortably during the testing. Do not flex your muscles.
Disabled people who find standing for minutes a little bit difficult can get support from the back or side. In
this case, there should be no skin-to-skin contact between the supporter and the patient. Testing is
impossible with an amputee who has a thumb, an arm or a leg amputated.
Summary of Contents for InBody 720
Page 1: ...User s Manual...
Page 44: ...1 Setup Menu 2 Quick Setup 3 DATABASE 4 Modification Example Chapter 3 Setup Establishment...
Page 59: ...1 Consumables 2 Basic Equipment 3 Options Chapter 5 Consumables...
Page 71: ......
Page 72: ...InBody the product of great technology Experience its speciality...
Page 76: ...www biospace co kr...