Indication
Error description
Error: blood sample Error
An insufficient amount of sample volume was applied in the test
panel.
Solution
Discard this test panel and test again using a new test panel. When
UFTUBHBJOBQQMZUIFQSPQFSWPMVNF BCPVUV-PGTBNQMFNJYUVSF
Indication
Error description
Error: Expired test kit
The test kits are expired.
Solution
Discard this test kit and test again using a new test kit does not past
expiration date.
Indication
Error description
Error: Temperature Error Unknown barcode
If the environmental temperature is above or below the operating
range of a analyzer, a thermo analyzer icon will appear on the display.
Solution
Move to an area at 15-32°C(59-90°F), and perform a test. Do not
artificially heat or cool the analyzer.
Indication
Error description
Error: Communication Error
The communication between analyzer and computer or printer is
failed.
Solution
Connect again between analyzer and external device.
If the error continues turn off an analyzer. Then turn on the analyzer
again. If there is still error massage, please contact SD Biosensor, Inc.