2. Preparing a measurement
SD A1cCare™ testing system
t
SD A1cCare™ Analyzer
t
SD A1cCare™ Test Kit
t
-BODFUTBOE-BODJOHEFWJDF
t
V-V-QJQFUUF 5ZQF"
t
Alcohol swab, if required
Precaution
Always
t
Read the test kit insert.
t
Operate the analyzer at the acceptable test-specific temperatures 15~32°C
(59~89°F). Place the analyzer on a level surface.
t
Keep the guideline for cleanliness.
t
Make sure that all display elements are displayed.
Never
t
Never Use test kits beyond the expiration printed on kit because they may cause
inaccurate results.
t
Never Touch or remove the test kit during actual measurement.
t
Never Subject the analyzer to sudden movements during a measurement.
t
Never Reuse the test kit applied either blood or control solution. Store the analyzer
and test kit at extreme temperatures.
t
Never Store the analyzer and test kits under humid or damp conditions without
suitable protection.
t
Only use SD A1cCare
TM
test kits. Using other test kits with this analyzer can cause inaccurate
results.
Not following these precautions can lead to inaccurate results.