10
5. Calibration
Attention!
Calibration is available for models of version V.6AW (PST-
60HL) and V.4AW (PST-60HL-4).
5.1.
The device is pre-calibrated at the factory (calibrating coefficient is 1.000) for
operation with temperatures, measured by a sensor, installed in the heating
block.
5.2.
To enter the calibration coefficient, hold the
Stop
key (fig. 1/8) pressed for
more than 8 s to activate calibration mode. The calibration coefficient will be
shown on the display (fig. 2/1).
5.3.
Restoring factory settings.
Temp
keys
(fig. 1/6) as shown on Fig. 2/1 to restore the factory settings. Press the
Run
key (fig. 1/7) once to save the changes and exit the calibration mode.
Note.
Coefficient value changes are recommended after the unit has
5.4.1.
Calibration procedure.
spectrocells, placed into the block sockets.
5.4.2.
Fig.2 Control panel in calibration mode
5.4.3. After the unit reaches the set temperature (when the set and current
temperature readings equal) leave the unit for 30 minutes for thermal
stabilization.
5.4.4.
s actual temperature is 40
correction.
5.4.5. Hold
Stop
key (fig. 1/8) pressed for more than 8 s to activate calibration mode.
The following parameters will be shown on the display (fig. 2).