5.  Calibration 

 

Attention!

 

Calibration  is  available  for  models  of  version  V.6AW  (PST-
60HL) and V.4AW (PST-60HL-4). 

5.1. 

The device is pre-calibrated at the factory (calibrating coefficient is 1.000) for 
operation with temperatures, measured by a sensor, installed in the heating 
block.  

5.2. 

To  enter  the calibration coefficient, hold the 

Stop

 key (fig. 1/8) pressed  for 

more than 8 s to activate calibration mode. The calibration coefficient will be 
shown on the display (fig. 2/1). 

5.3. 

Restoring factory settings.

 

Temp 

keys 

(fig. 1/6) as shown on Fig. 2/1 to restore the factory settings. Press the 

Run

 

key (fig. 1/7) once to save the changes and exit the calibration mode. 

 

Note.

 

Coefficient value changes are recommended after the unit has 

 

5.4.1.  

Calibration  procedure.

 

spectrocells, placed into the block sockets. 

5.4.2.  

 

 
 
 

 

 
 

Fig.2 Control panel in calibration mode 

5.4.3.  After  the  unit  reaches  the  set  temperature  (when  the  set  and  current 

temperature  readings  equal)  leave  the  unit  for  30  minutes  for  thermal 
stabilization. 

5.4.4.  

s actual temperature is 40

correction. 

5.4.5.  Hold 

Stop

 key (fig. 1/8) pressed for more than 8 s to activate calibration mode. 

The following parameters will be shown on the display (fig. 2). 

 

 

DOMINIQUE DUTSCHER SAS

Summary of Contents for PST-60HL

Page 1: ...D O M I N I Q U E D U T S C H E R S A S ...

Page 2: ...2 D O M I N I Q U E D U T S C H E R S A S ...

Page 3: ...Precautions 4 2 General information 6 3 Getting started 7 4 Operation 8 5 Calibration 10 6 Specifications 12 7 Maintenance 13 8 Warranty and Claims 14 9 Declaration of Conformity 15 D O M I N I Q U E D U T S C H E R S A S ...

Page 4: ...fications to the design of the unit ELECTRICAL SAFETY Connect only to the external power supply unit with voltage corresponding to that on the serial number label Use only the external power supply unit provided with this product Ensure that the switch and external power supply are easily accessible during use Do not plug the unit into an ungrounded power socket and do not use an ungrounded extens...

Page 5: ...possible operation of the unit in specific atmospheres Do not operate the unit if it is faulty or has been installed incorrectly Do not fill in the microplates after they have been inserted in the unit Do not use outside laboratory rooms Do not check the temperature by touch Use a thermometer BIOLOGICAL SAFETY hazardous material is spilled on or penetrates into the equipment D O M I N I Q U E D U ...

Page 6: ... 60HL PST 60HL 4 Plate Shaker Thermostat provides soft or intensive sample shaking rotation speed regulation stabilization and indication even shaking amplitude throughout Shaker Thermostat platform required operation time setting and indication automatic stopping of the platform movement after the set time expires current operation time indication setting and indication of the required temperatur...

Page 7: ...ST 60HL PST 60HL 4 Plate Shaker Thermostat 1 pce spare rubber belt 2 pcs external power supply 1 pce power cord 1 pce Operating manual Certificate 1 copy 3 3 Set up place the unit upon even horizontal non flammable surface away from any flammable materials not less than 30 cm remove protective film from the display plug the external power supply unit into the socket at the rear side of the unit an...

Page 8: ...eadings in the upper line of the display Set while setting the necessary parameters Pressing the key for more than 3 s will increase the increment 4 3 1 Setting time TIME he required working time interval in hours and minutes increment 1 min 4 3 2 Setting speed RPM uired speed increment 10 RPM 4 3 3 Caution The platform heating can be turned off only by setting the set point It can be used in cold...

Page 9: ...s operation of the Thermo shaker timer indicator will start counting up the time interval in the lower line Actual until the RPM RUN STOP key is pressed again Please note The platform temperature will be constantly maintained in accordance with the set temperature This allows using the device again without pre heating 4 6 The timer can be reset during operation if required Press the TIME RUN STOP ...

Page 10: ...ig 1 6 as shown on Fig 2 1 to restore the factory settings Press the Run key fig 1 7 once to save the changes and exit the calibration mode Note Coefficient value changes are recommended after the unit has 5 4 1 Calibration procedure spectrocells placed into the block sockets 5 4 2 Fig 2 Control panel in calibration mode 5 4 3 After the unit reaches the set temperature when the set and current tem...

Page 11: ... correct temperature through all the operation range Note Coefficient value changes are recommended after the unit has 5 4 7 After finishing the calibration press the Run key fig 1 7 once to save the changes and exit the calibration mode 5 4 8 The display will show calibrated temperature as shown on fig 4 1 and the unit will continue thermal stabilization according to the previously set temperatur...

Page 12: ...n operation sessions not less than 8 hours Time setting resolution 1 min Max height of microtest plate 18 mm External power supply input AC 100 240 V 50 60 Hz output DC 12 V PST 60HL PST 60HL 4 Current power consumption 12 V DC 3 3 A 40 W 12 V DC 4 15 A 50 W Number of microplates 2 4 Platform dimensions 250 x 150 mm 210 x 290 mm Dimensions 270x260x125 mm 380x390x140 mm Weight 6 1 kg 8 8 kg Replace...

Page 13: ...ation time To replace the belt Disconnect the external power supply unit from the device Remove 4 fixation screws on the underside and remove the bottom plate Replace the rubber belt fig 5 Reassemble the device Fig 5 Rubber belt replacement 7 5 Error codes in case of a defect Following error codes are shown in the lower right corner of the display fig 1 accompanied by a sound signal every 8 s Pres...

Page 14: ...ranty 8 3 Warranty covers only the units transported in the original package 8 4 If any manufacturing defects are discovered by the Customer an unsatisfactory equipment claim shall be compiled certified and sent to the local distributor address Please visit www biosan lv Technical support section to obtain the claim form 8 5 The following information will be required in the event that warranty or ...

Page 15: ...15 9 Declaration of Conformity Edition 4 6 01 April 2016 D O M I N I Q U E D U T S C H E R S A S ...

Page 16: ...D O M I N I Q U E D U T S C H E R S A S ...

Reviews: