8
4. Operation
Recommendations during operation
For proper mixing, it is not recommended to fill microplates for more than 75%
of their nominal volume.
Seal the microplates with an appropriate adhesive film to avoid spilling the
sample.
Fig.1 Control panel
4.1.
Connect the external power supply unit to a grounded power socket and set
the power switch located on the rear panel of the unit to position
I
(ON).
4.2.
The display will turn on. Upper line (
Set
) shows time, speed and temperature
set earlier. Lower line (
Actual
) shows current readings of the same
according to the temperature set in the upper line). The time of temperature
stabilisation depends on the initial temperature.
4.3.
Setting the parameters. Use the readings in the upper line of the display (
Set
),
while setting the necessary parameters. Pressing the key for more than 3 s
will increase the increment.
4.3.1.
Setting time (TIME)
he required
working time interval in hours and minutes (increment 1 min).
4.3.2.
Setting speed (RPM).
uired
speed (increment 10 RPM).
4.3.3.
Caution!
The platform heating can be turned off only by setting the
-
- set point).It can be used in cold rooms as a mixing
device without thermal regulation in this mode.
4.4.
Program execution
. After the thermal stabilisation of the unit (when the set
and current temperature readings become the same):
4.4.1. Microplate fixation:
PST-60HL:
Place microplates on the platform and fix it with the special
push-down clip by pressing it against the plate covers.
PST-60HL-4: Unscrew the fixation screws. Place microplates on the
platform and fix them by tightening the fixation screws.