4.  Operation  

Recommendations during operation 

 

For proper mixing, it is not recommended to fill microplates for more than 75% 

of their nominal volume.  

 

Seal the  microplates  with  an  appropriate  adhesive  film  to  avoid  spilling  the 

sample.  

 

 

Fig.1 Control panel 

4.1. 

Connect the external power supply unit to a grounded power socket and set 
the power switch located on the rear panel of the unit to position 

I

 (ON). 

4.2. 

The display will turn on. Upper line (

Set

) shows time, speed and temperature 

set  earlier.  Lower  line  (

Actual

)  shows  current  readings  of  the  same 

according to the temperature set in the upper line). The time of temperature 
stabilisation depends on the initial temperature. 

4.3. 

Setting the parameters. Use the readings in the upper line of the display (

Set

), 

while setting the necessary parameters. Pressing the key for more than 3 s 
will increase the increment. 

4.3.1. 

Setting  time  (TIME)

he  required 

working time interval in hours and minutes (increment 1 min).  

4.3.2. 

Setting speed (RPM).

 

uired 

speed (increment 10 RPM). 

4.3.3. 

 

 

 

Caution!

 

The  platform  heating  can  be  turned  off  only  by  setting  the 

-  

-  set point).It can be used in cold rooms as a mixing 

device without thermal regulation in this mode. 

4.4. 

Program execution

. After the thermal stabilisation of the unit (when the set 

and current temperature readings become the same): 

4.4.1.  Microplate fixation:  
 

PST-60HL: 

Place microplates  on the platform and fix it  with  the special 
push-down clip by pressing it against the plate covers. 

 

PST-60HL-4:  Unscrew  the  fixation  screws.  Place  microplates  on  the 

platform and fix them by tightening the fixation screws. 

DOMINIQUE DUTSCHER SAS

Summary of Contents for PST-60HL

Page 1: ...D O M I N I Q U E D U T S C H E R S A S ...

Page 2: ...2 D O M I N I Q U E D U T S C H E R S A S ...

Page 3: ...Precautions 4 2 General information 6 3 Getting started 7 4 Operation 8 5 Calibration 10 6 Specifications 12 7 Maintenance 13 8 Warranty and Claims 14 9 Declaration of Conformity 15 D O M I N I Q U E D U T S C H E R S A S ...

Page 4: ...fications to the design of the unit ELECTRICAL SAFETY Connect only to the external power supply unit with voltage corresponding to that on the serial number label Use only the external power supply unit provided with this product Ensure that the switch and external power supply are easily accessible during use Do not plug the unit into an ungrounded power socket and do not use an ungrounded extens...

Page 5: ...possible operation of the unit in specific atmospheres Do not operate the unit if it is faulty or has been installed incorrectly Do not fill in the microplates after they have been inserted in the unit Do not use outside laboratory rooms Do not check the temperature by touch Use a thermometer BIOLOGICAL SAFETY hazardous material is spilled on or penetrates into the equipment D O M I N I Q U E D U ...

Page 6: ... 60HL PST 60HL 4 Plate Shaker Thermostat provides soft or intensive sample shaking rotation speed regulation stabilization and indication even shaking amplitude throughout Shaker Thermostat platform required operation time setting and indication automatic stopping of the platform movement after the set time expires current operation time indication setting and indication of the required temperatur...

Page 7: ...ST 60HL PST 60HL 4 Plate Shaker Thermostat 1 pce spare rubber belt 2 pcs external power supply 1 pce power cord 1 pce Operating manual Certificate 1 copy 3 3 Set up place the unit upon even horizontal non flammable surface away from any flammable materials not less than 30 cm remove protective film from the display plug the external power supply unit into the socket at the rear side of the unit an...

Page 8: ...eadings in the upper line of the display Set while setting the necessary parameters Pressing the key for more than 3 s will increase the increment 4 3 1 Setting time TIME he required working time interval in hours and minutes increment 1 min 4 3 2 Setting speed RPM uired speed increment 10 RPM 4 3 3 Caution The platform heating can be turned off only by setting the set point It can be used in cold...

Page 9: ...s operation of the Thermo shaker timer indicator will start counting up the time interval in the lower line Actual until the RPM RUN STOP key is pressed again Please note The platform temperature will be constantly maintained in accordance with the set temperature This allows using the device again without pre heating 4 6 The timer can be reset during operation if required Press the TIME RUN STOP ...

Page 10: ...ig 1 6 as shown on Fig 2 1 to restore the factory settings Press the Run key fig 1 7 once to save the changes and exit the calibration mode Note Coefficient value changes are recommended after the unit has 5 4 1 Calibration procedure spectrocells placed into the block sockets 5 4 2 Fig 2 Control panel in calibration mode 5 4 3 After the unit reaches the set temperature when the set and current tem...

Page 11: ... correct temperature through all the operation range Note Coefficient value changes are recommended after the unit has 5 4 7 After finishing the calibration press the Run key fig 1 7 once to save the changes and exit the calibration mode 5 4 8 The display will show calibrated temperature as shown on fig 4 1 and the unit will continue thermal stabilization according to the previously set temperatur...

Page 12: ...n operation sessions not less than 8 hours Time setting resolution 1 min Max height of microtest plate 18 mm External power supply input AC 100 240 V 50 60 Hz output DC 12 V PST 60HL PST 60HL 4 Current power consumption 12 V DC 3 3 A 40 W 12 V DC 4 15 A 50 W Number of microplates 2 4 Platform dimensions 250 x 150 mm 210 x 290 mm Dimensions 270x260x125 mm 380x390x140 mm Weight 6 1 kg 8 8 kg Replace...

Page 13: ...ation time To replace the belt Disconnect the external power supply unit from the device Remove 4 fixation screws on the underside and remove the bottom plate Replace the rubber belt fig 5 Reassemble the device Fig 5 Rubber belt replacement 7 5 Error codes in case of a defect Following error codes are shown in the lower right corner of the display fig 1 accompanied by a sound signal every 8 s Pres...

Page 14: ...ranty 8 3 Warranty covers only the units transported in the original package 8 4 If any manufacturing defects are discovered by the Customer an unsatisfactory equipment claim shall be compiled certified and sent to the local distributor address Please visit www biosan lv Technical support section to obtain the claim form 8 5 The following information will be required in the event that warranty or ...

Page 15: ...15 9 Declaration of Conformity Edition 4 6 01 April 2016 D O M I N I Q U E D U T S C H E R S A S ...

Page 16: ...D O M I N I Q U E D U T S C H E R S A S ...

Reviews: