8
6. Specifications
The unit is designed for operation in cold rooms, incubators (except CO
2
incubators)
and closed laboratory rooms at ambient temperature from +4°C to +40°C in a non-condens-
ing atmosphere and maximum relative humidity 80% for temperatures up to 31°C decreasing
l
inearly to 50% relative humidity at 40°C.
Biosan is committed to a continuous programme of improvement and reserves the
right to alter design and specifications of the equipment without additional notice.
MS-3000
MMS-3000
Speed control range
0
–3000 rpm
Maximum stirring volume (H
2
O)
5 L
20 L
Working surface dimensions
110x110 mm
Ø160 mm
Working surface material
Stainless steel
SR-1 detachable stand size
—
Ø8x320 mm
Maximum length of magnetic
stirring element
50 mm
70 mm
Maximum stirred liquid viscosity
Up to 1170 mPa*s
Maximum continuous operation time
1
24 h
Working current / Power consumption
220 mA / 2.6 W
250 mA / 3 W
External power supply
in AC 100-240 V, 50/60 Hz, out DC 12 V
Weight
2
0.8 kg
1.5 kg
Dimensions
120x150x65 mm
185x230x75 mm
Table 1. MS-3000 and MMS-3000 maximal speed in RPM, depending on H
2
O volume and mag-
netic element length
Magnetic element length, mm
MS-3000
MMS-3000
2 L
5 L
2 L
5 L
15 L
20 L
25
3000
2400
3000
2800
2650
2600
50
700
700
1700
1500
1300
1250
70
–
–
620
530
440
360
1
Maintain at least 1 hour long pause between prolonged continuous operations
2
Accurate within ± 10%