background image

 

42

EMC Information 

The  BionicGym  complies  with  ANSI  AAMI  IEC  60601-1-2:2014  standard  for 

Electromagnetic Compatibility.

Electromagnetic Emissions 

The  BionicGym  is  intended  for  use  in  the  electromagnetic  environment  specified 

below.  The user should ensure that it is used in such an environment. 

Emissions Test 

Compliance  Guidance 

IEC 61000-3-2:2014  Complies 

The  BionicGym  is  suitable  for  use  in  all 

establishments,  including  domestic  and  those 

directly  connected  to  the  public  low-voltage 

power supply network that supplies Harmonic 

emissions  buildings  used  for  domestic 

purposes. 

IEC 61000-3-

3:2013+A1:2017 

Complies 

Electromagnetic Immunity 

Standard 

Item 

IEC 60601-1-2 

Test Levels 

BionicGym Test 

Levels  

IEC 61000-4-2:2008  ESD 

± 8 kV contact;  

± 2 kV, ± 4 kV,  

± 8 kV, ± 15 kV air 

± 2 kV, ± 4 kV, ± 8 

kV contact;  

± 2 kV,± 4 kV,  

± 8 kV, ± 15 kV air 

IEC 61000-4-

3:2006+A1:2007+A

2:2010 

RS 

10 V/m 80 MHz – 2.7 GHz 80% 

AM at 1 kHz  

10 V/m 80 MHz – 

5.785 GHz 80% 

AM at 1 kHz  

IEC 61000-4-4:2012  EFT 

± 2 kV 100 kHz repetition 

frequency  

± 2 kV 100 kHz 

repetition 

frequency 

IEC 61000-4-5:2014  Surge  ± 0.5 kV, ± 1 kV 

± 0.5 kV, ± 1 kV 

IEC 61000-4-6:2013  CS 

3V 0.15 MHz – 80 MHz  

6V in ISM and amateur bands 

between 0.15 MHz and 80 

MHz 80% AM at 1 kHz  

10V  0.15 MHz – 

80  MHz 80% AM 

at 1 kHz  

IEC 61000-4-8:2009  PFMF  30 A/m 

50 Hz or 60 Hz 

± 0.5 kV, 

± 1 kV 

IEC 61000-4- 11:2004 

 

Voltage Dips: 

43

Voltage dips & voltage 

variations

 

1) 0% UT; 0.5 cycle at 0º. 45º, 90º, 135º, 225º,

270º, 315º

2) 0% UT; 1 cycle; Single phase at 0º

3) 70% UT; 25/30 cycles;

Single phase at 0º

Voltage Interruptions: 0% UT; 250/300 cycle;

Radio Frequency Wireless Information 

This device complies with Part 15 of the FCC Rules. Operation is subject to the following 

two conditions: (1) this device may not cause harmful interference, and (2) this device 

must accept any interference received, including interference that may cause undesired 

operation. 

FCC ID: 2ARJQBG01 

Radiated output power (RF): 2.4 GHz, 3 mW Max. 

Recommended Separation Distances 

The  BionicGym  contains  a  Bluetooth®  Radio  Module  and  it  is  intended  for  use  in  an 

electromagnetic environment in which radiated RF disturbances are controlled.  The user 

of  the  BionicGym  can  help  prevent  electromagnetic  interference  by  maintaining  a 

minimum  distance  between  portable  and  mobile  RF  communications  equipment 

(transmitters) and  the BionicGym as recommended  below,  according to  the maximum 

output power of the communications equipment. 

Recommended separation distances between portable and mobile RF communication 

equipment and the BionicGym. 

Rated maximum 

output power of 

transmitter W 

Separation distance according to frequency of transmitter m 

150 KHz to 

80 MHz 

80 MHz to 

800 MHz 

800 MHz to 

2.5 GHz 

d = 1.17 √ P 

d = 1.17 √ P 

d = 2.33 √ P 

0.01 

0.12 

0.12 

0.23 

0.1 

0.37 

0.37 

0.75 

1.17 

1.17 

2.33 

10 

3.70 

3.70 

7.36 

100 

11.70 

11.70 

23.50 

NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.

NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by

absorption and reflection from structures, objects and people.

42

43

For transmitters rated at a maximum output power not listed above, the recommended 

separation distance d in metres (m) can be estimated using the equation applicable to 

the frequency  of  the  transmitter,  where  P  is  the  maximum  output  power  rating  of 

the transmitter in watts (w) according to the transmitter manufacturer. 

NOTE 1

:  At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.

NOTE 2

:  These guidelines may not apply in all situations. Electromagnetic propagation is a

ff

ected by

absorption and r

eflec

tion from structures, objects and people.

Summary of Contents for BG-300-02

Page 1: ...This page is not for printing USER MANUAL Draft v 0 87_LC_EA_MB_HC_25 01 2019 User Manual Rev03...

Page 2: ...ce 10 6 In the BionicGym box 11 7 Getting Started 14 8 Setup and First Use 15 9 Download and install the App 22 10 Auto Stop 29 11 Workout Sessions 29 12 Standard Workout Programs 29 13 Zones 31 14 Bi...

Page 3: ...uccess Sincerely Dr Louis Crowe Inventor CEO 2 3 3 2 Overview Exercise happily safely and comfortably Congratulations You are one of the first owners of BionicGym a wearable exercise machine EXERCISE...

Page 4: ...or Stop a session via the app In an emergency you could also remove the control unit from the pocket Use the app to unpause or double press the orange button 4 5 5 3 Tips to get the most out of Bioni...

Page 5: ...including clots DVTs You should ask your doctor if it is safe for you to exercise any exercise including BionicGym especially if you are injured over 40 smoke are obese or diabetic have heart disease...

Page 6: ...inery or during any activity in which involuntary muscle contractions may put the user at undue risk of injury 8 9 Adverse Reactions Skin irritation and burns beneath the electrodes have been reported...

Page 7: ...th the European Space Agency I ve suffered for and been elated with the science Please visit our website and read the peer reviewed scientific publications 10 11 6 In the BionicGym box DEVICE ACCESSOR...

Page 8: ...ps to the end of the control unit with magnets If not pre attached the charger plugs into the bottom of the cradle Charger Adaptor The charger is supplied with an adaptor for your country Align and tw...

Page 9: ...ully charged 14 15 8 Setup and First Use Have you 1 Reviewed the safety information And checked with your Doctor if you have any doubts or medical issues 2 Watch the training videos first time use vid...

Page 10: ...on the leg FRONT VIEW ELECTRODES IN THE MID LINE 16 17 REAR VIEW ELECTRODES ALSO MID LINE This is where the electrodes should end up MID LINE You will stick them onto the wraps so that they will be c...

Page 11: ...t the wrap on and check the electrode position s Reposition the electrode s if required Note you can also stretch the wrap as you put it on to make small adjustments For larger legs both small and lar...

Page 12: ...n facing out 21 The wire auxiliary cable connects the right and left wraps Push it in firmly into the sockets on both side The jack sockets are located in pockets You can keep the wire tidy by running...

Page 13: ...iOS Apple Android Google iOS 8 Lollipop 5 0 5 1 1 iOS 9 Marshmallow 6 0 6 01 iOS 10 Nougat 7 0 7 1 1 23 Registering the device To register you will need your phone pad connected to the internet You w...

Page 14: ...d itself OFF if it was not paired within a couple of minutes Make sure the unit is ON LED is flashes Blue if trying to pair to the app and CHARGED the LED may be flashing Red or not switched on if the...

Page 15: ...r resume it on the App The circle changes from white to orange as you progress through the session If the unit loses connection with the App it may need to reconnect follow the app instructions Even w...

Page 16: ...ess you interact with the App your phone device every few minutes Note this is not available in all Apps 11 Workout Sessions These may be updated from time to time From June 2021 programs will be desc...

Page 17: ...n out in public places You may not be able to reach the same intensities as the other programs and may find it more fatiguing For most the measured cardio aerobic effect will be lower than our standar...

Page 18: ...who want to train hard As with the Standard programs it is important to find which works best for you As a general rule the higher the intensity you can go on a program the better it will be for you I...

Page 19: ...before use Sometimes if you leave the wraps on for a while before starting the stimulation the gel surface dries out and you need to wet the skin before continuing If you have a small cut or blemish t...

Page 20: ...from the grid side of the electrodes Are the wraps on the legs and all the wiring firmly connected and the unit cleanly docked in the pocket Switch unit on and off Is the battery charging full If it...

Page 21: ...s not working correctly Do not use in the meantime 38 Understanding the LED Lights LED Light State Blue Flashing Bluetooth trying to pair with the app Bluetooth connection lost during workout for a pe...

Page 22: ...n objects 12 5 mm Keep the unit dry at all times Neurostimulation All electrical specifications are supplied for a load impedance from 500 to 1 000 Typical session duration per workout program 1 hour...

Page 23: ...nterference and 2 this device must accept any interference received including interference that may cause undesired operation FCC ID 2ARJQBG01 Radiated output power RF 2 4 GHz 3 mW Max Recommended Sep...

Page 24: ...aution and Precaution to note CE Mark Federal Communications Commission Manufactured By Use By Batch Number Serial Number EC Authorized Representative Temperature Range Humidity Range Type BF Applied...

Page 25: ...the BionicGym name logos etc are trademarks of Medical Currents Ltd The terms of use specify that no attempt to be made to reverse engineer or copy any aspect of the device outputs or associated app...

Page 26: ...nterest in BionicGym especially to our backers on Indiegogo that have made this possible TOP TIP Make BionicGym a habit Always use it say after lunch standing at your desk for an hour or watching your...

Page 27: ...www BionicGym com...

Reviews: