42
EMC Information
The BionicGym complies with ANSI AAMI IEC 60601-1-2:2014 standard for
Electromagnetic Compatibility.
Electromagnetic Emissions
The BionicGym is intended for use in the electromagnetic environment specified
below. The user should ensure that it is used in such an environment.
Emissions Test
Compliance Guidance
IEC 61000-3-2:2014 Complies
The BionicGym is suitable for use in all
establishments, including domestic and those
directly connected to the public low-voltage
power supply network that supplies Harmonic
emissions buildings used for domestic
purposes.
IEC 61000-3-
3:2013+A1:2017
Complies
Electromagnetic Immunity
Standard
Item
IEC 60601-1-2
Test Levels
BionicGym Test
Levels
IEC 61000-4-2:2008 ESD
± 8 kV contact;
± 2 kV, ± 4 kV,
± 8 kV, ± 15 kV air
± 2 kV, ± 4 kV, ± 8
kV contact;
± 2 kV,± 4 kV,
± 8 kV, ± 15 kV air
IEC 61000-4-
3:2006+A1:2007+A
2:2010
RS
10 V/m 80 MHz – 2.7 GHz 80%
AM at 1 kHz
10 V/m 80 MHz –
5.785 GHz 80%
AM at 1 kHz
IEC 61000-4-4:2012 EFT
± 2 kV 100 kHz repetition
frequency
± 2 kV 100 kHz
repetition
frequency
IEC 61000-4-5:2014 Surge ± 0.5 kV, ± 1 kV
± 0.5 kV, ± 1 kV
IEC 61000-4-6:2013 CS
3V 0.15 MHz – 80 MHz
6V in ISM and amateur bands
between 0.15 MHz and 80
MHz 80% AM at 1 kHz
10V 0.15 MHz –
80 MHz 80% AM
at 1 kHz
IEC 61000-4-8:2009 PFMF 30 A/m
50 Hz or 60 Hz
± 0.5 kV,
± 1 kV
IEC 61000-4- 11:2004
Voltage Dips:
43
Voltage dips & voltage
variations
1) 0% UT; 0.5 cycle at 0º. 45º, 90º, 135º, 225º,
270º, 315º
2) 0% UT; 1 cycle; Single phase at 0º
3) 70% UT; 25/30 cycles;
Single phase at 0º
Voltage Interruptions: 0% UT; 250/300 cycle;
Radio Frequency Wireless Information
This device complies with Part 15 of the FCC Rules. Operation is subject to the following
two conditions: (1) this device may not cause harmful interference, and (2) this device
must accept any interference received, including interference that may cause undesired
operation.
FCC ID: 2ARJQBG01
Radiated output power (RF): 2.4 GHz, 3 mW Max.
Recommended Separation Distances
The BionicGym contains a Bluetooth® Radio Module and it is intended for use in an
electromagnetic environment in which radiated RF disturbances are controlled. The user
of the BionicGym can help prevent electromagnetic interference by maintaining a
minimum distance between portable and mobile RF communications equipment
(transmitters) and the BionicGym as recommended below, according to the maximum
output power of the communications equipment.
Recommended separation distances between portable and mobile RF communication
equipment and the BionicGym.
Rated maximum
output power of
transmitter W
Separation distance according to frequency of transmitter m
150 KHz to
80 MHz
80 MHz to
800 MHz
800 MHz to
2.5 GHz
d = 1.17 √ P
d = 1.17 √ P
d = 2.33 √ P
0.01
0.12
0.12
0.23
0.1
0.37
0.37
0.75
1
1.17
1.17
2.33
10
3.70
3.70
7.36
100
11.70
11.70
23.50
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
42
43
For transmitters rated at a maximum output power not listed above, the recommended
separation distance d in metres (m) can be estimated using the equation applicable to
the frequency of the transmitter, where P is the maximum output power rating of
the transmitter in watts (w) according to the transmitter manufacturer.
NOTE 1
: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2
: These guidelines may not apply in all situations. Electromagnetic propagation is a
ff
ected by
absorption and r
eflec
tion from structures, objects and people.