Ver. 1.52 4.Operation 32
[ Description ]
Setup Menu
Description
Max SpO
₂
(UPPER)
Max SpO
₂
for alarm can be 41-100%. It can be adjusted by 1%
( Menu )
( Alarm setup Screen )
( Adjust Max Limit )
Min SpO
₂
(LOWER)
Min SpO
₂
for alarm can be 40~99%. It can be adjusted by 1%
( Menu ) ( Alarm Setup )
( Move to Min Limit ) ( Adjust Min Limit )
NOTE
: limit must be smaller than Max limit.
Max PULSE RATE
(UPPER)
The max BPM for alarm can be 31bpm~250bpm, and it can be adjusted by
10bpm.
Min PULSE RATE
(LOWER)
The min BPM for alarm can be 30bpm~249bpm, and it can be adjusted by
1bpm
NOTE
: The Min limit must be smaller than the Max limit.
LEVEL
(% SpO
₂
, BPM)
This menu allows the user to set the alarm level. (HIGH, MEDIUM, LOW)
VOLUME
This menu allows the user to set the alarm volume (5step.)
WARNING
: Any alarm levels can be turned off, including lethal alarms.
Parameters configured to use All Off do not have any visual or audio alarm
indicators
Any alarm levels can be turned off, including lethal alarms. Parameters configured to use All Off do
not have any visual or audio alarm indicators.
Any alarm levels can be turned off, including lethal alarms. Parameters configured to use All Off do
not have any visual or audio alarm indicators.