151
BM5Vet Pro
User’s Manual
8. If the cuvette pressure value changes more, troubleshoot tubing and fittings. If the
problem persists, return the analyzer to PHASEIN.
Note:
In step 5, if the cuvette pressure is less than 15 kPa below atmospheric
pressure, repeat steps 1 to 3 blocking the exhaust port quicker.
Gas span check
Confirm the gas measurements values by performing the following sequence
1. Connect a new Nomoline sampling line to the ISA gas analyzer.
2. Warm up for at least 1 min.
3. Connect the calibration gas to the ISA gas analyzer. The calibration gas shall have
an accuracy of at least ±
0.03 vol% or ±
(0.02 vol% +0.1% of reading), whichever
is greater.
4. Supply the calibration gas.
5. Check that the displayed gas concentration readings correspond to the calibration
gas values.
6. The gas readings shall be within the following ranges:
a. CO
2
± (
0.3 vol% + 4% of reading)
b. N
2
O
± (
2 vol% + 5% of reading)
c. Agents ±
(0.2 vol% + 10% of reading)
d. O
2
± (
2 vol% + 2% of reading)
7. If the gas readings are outside the specified range, perform a gas span calibration for
the failing gases. If failing on Agents, the span calibration shall always be performed
using DES.
Gas span calibration
Only
perform the gas span calibration if the gas span check fails repeatedly.
Before performing the gas span calibration, ensure that the SET O2 and SET N2O
Summary of Contents for BM5VET
Page 22: ...22 BM5Vet Pro User s Manual Left Side 1 Printer...
Page 68: ...68 BM5Vet Pro User s Manual ECG Lead 5 Positions of 5 Lead Placement RL V LL LA RA...
Page 155: ...155 BM5Vet Pro User s Manual Accessories Bionet Dual Gas Module...
Page 195: ...195 BM5Vet Pro User s Manual ACCVENT COUPLET IRREGULAR PAUSE RONT VBRADY SHORTRUN PVC...
Page 219: ...219 BM5Vet Pro User s Manual SYSTEM LOWBATTERY...
Page 225: ...225 BM5Vet Pro User s Manual less than minus number percent plus or minus...