186
BM5Vet Pro
User’s Manual
a-
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telepho
nes and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot
be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed
RF transmitters, an electromagnetic site survey should be considered. If the measured field strengt
h outside the shielded location in which the EUT is used exceeds 3V/m, the EUT should be obser
ved to verify normal operation.
If abnormal performance is observed, additional measures may be necessary, such as relocating th
e EUT or using a shielded location with a higher RF shielding effectiveness and filter attenuation.
Note
For Type A Professional ME Equipment intended for use in domestic establishment instructions for
use includes a warning:
This ME equipment is intended for use by professional healthcare personnel only.
Warning
Portable RF communications equipment (including peripherals such as antenna cables and external
antennas) should be used no closer than 30 cm (12 inches) to any part of the [ME EQUIPMENT or ME
SYSTEM], including cables specified by the manufacturer
Warning
Use of accessories and cables other than those specified or provided by the manufacturer of this
equipment could result in increased electromagnetic emissions or decreased electromagnetic
immunity of this equipment and result in improper operation
Summary of Contents for BM5VET
Page 22: ...22 BM5Vet Pro User s Manual Left Side 1 Printer...
Page 68: ...68 BM5Vet Pro User s Manual ECG Lead 5 Positions of 5 Lead Placement RL V LL LA RA...
Page 155: ...155 BM5Vet Pro User s Manual Accessories Bionet Dual Gas Module...
Page 195: ...195 BM5Vet Pro User s Manual ACCVENT COUPLET IRREGULAR PAUSE RONT VBRADY SHORTRUN PVC...
Page 219: ...219 BM5Vet Pro User s Manual SYSTEM LOWBATTERY...
Page 225: ...225 BM5Vet Pro User s Manual less than minus number percent plus or minus...