detected and has non-ventricular beats before and after.
R ON T:
Occurs when a ventricular complex is detected within the repolarization period of a non-
ventricular beat.
TRIGEMINY:
Occurs when two or more trigeminal cycles (a ventricular beat followed by two non-
ventricular beats) are detected.
V BRADY:
Ventricular bradycardia occurs when a run of three or more ventricular beats is
detected with an average heart rate that is less than or equal to 50 beats per minute.
NOTE
The Brady limit matches the low heart rate limit. If the low heart rate limit is changed, the
Brady limit changes.
NOTE
The Tachy limit matches the high heart rate limit. If the high heart rate limit is changed, the
Tachy limit changes.
Arrhythmia Settings
A. ECG menu
MENU
Description
Available Settings
A-1.
Arrhythmia
Arrhythmia parameter alarm setup
menu.
A-1-1.
Arrhythmia Type
Arrhythmia diagnosis level setup menu.
OFF: Do not perform arrhythmia
diagnosis.
LETHAL: Performs the detection of
Asys, VTAC/VFIB, and VTAC at the
selected lead
FULL: Performs the detection of all 13
arrhythmia.
OFF, LETHAL, FULL
A-1-2.
Arrhythmia Alarm
Condition
Alarm setting menu by arrhythmia type
Summary of Contents for BM5VET
Page 22: ...22 BM5Vet Pro User s Manual Left Side 1 Printer...
Page 68: ...68 BM5Vet Pro User s Manual ECG Lead 5 Positions of 5 Lead Placement RL V LL LA RA...
Page 155: ...155 BM5Vet Pro User s Manual Accessories Bionet Dual Gas Module...
Page 195: ...195 BM5Vet Pro User s Manual ACCVENT COUPLET IRREGULAR PAUSE RONT VBRADY SHORTRUN PVC...
Page 219: ...219 BM5Vet Pro User s Manual SYSTEM LOWBATTERY...
Page 225: ...225 BM5Vet Pro User s Manual less than minus number percent plus or minus...