33
www.bioneer.com
BQ-042-101-12
Revision : 0 (2018-07-05)
ExiPrep
™
96 Lite
XI. User Maintenance
1.
Cleaning of the main body
Note
▪
Turn off the power of
ExiPrep
™
96 Lite and disconnect the power cord.
▪
Wait until the temperature of heat block drops back to room temperature.
(1)
Cleaning of the LCD Touch Screen
▪
Clean the LCD touch screen using the cleaning solution available in the market
▪
Be careful not to leave any scratch on the screen
(2)
Cleaning of the Door
▪
Clean the Door using the cleaning solution available in the market
▪
Be careful not to leave any scratch on the screen.
2.
Replacing the Fuse
This instrument uses two fuses.
▪
Tools required for the replacement
-
2 fuses with the specification of 250V, F6.3
AL, 5×20mm
-
1 flat-head screwdriver (small)
▪
Replacing the Fuse
1)
Turn off the power at the rear of
ExiPrep
™
96 Lite and disconnect the power cord.
Risk of electric shock
Serous electric shock may result during the operation of high voltage instrument and it may cause
physical injury or even death. To prevent the electric shock, disconnect the power cord at least one
minute before use and disconnect the power supply from the external power source.
2)
If you insert a flat-head screwdriver into the upper part of power inlet, the fuse cover can be opened.