VII. Removing the shipping bracket and approving power source
-
Clean up the test table that you want to place the instrument
-
Refer to this manual to check if all accessories are in included in the box
-
If the instrument does not work after removing the shipping bracket, and connecting power after
installation or initialization process is being completed, stop the operation immediately and contact us
or your dealer.
1.
Removing shipping bracket
1)
When you open the door, inside the instrument
you can see the bolts (A, B) and shipping
bracket(C) that holding the setup tray and
syringe in the front.
The shipping bracket must be removed before
approving the power source. Turning the power
source and operating the instrument without
first removing the shipping bracket may
damage the instrument.
2)
Remove two bolts (A, B) that are connected to
Buffer Cartridges Slides.
<Front View> <Rear View>
3)
When the bolts (A, B) are removed, the
shipping bracket(C) can be separated.
Store the shipping bracket and bolts removed
separately so that you can use it again when
relocating the instrument.