19
www.bioneer.com
BQ-042-101-12
Revision : 0 (2018-07-05)
ExiPrep
™
96 Lite
3.
Changes of Magnetic Rods
ExiPrep
™96 Lite can be used
with 96 wells and 24 wells depending on the user's samples and needs. It can
be used simply by replacing magnetic rods.
1)
After power on and initialization, open the
door.
2)
Click on the top of the ‘Plate’ button
.
Then the magnetic rod cover plate comes
down to insert the magnetic rod cover and
buffer cartridge layer 1 comes forward to
insert the cartridges.
3)
Remove the bracket at the top of the
magnetic rods plate by turning it to the left.
4)
Pulling the magnetic rods toward the front of
the instrument will remove the magnetic
rods.