30253 Rev A, Saphyr System Safety Guide for Saphyr P/N 60325 

Page 5 of 9 

Saphyr System Safety Guide

 

For Research Use Only. Not for use in diagnostic procedures. 

This guide provides important safety information pertaining to the installation, servicing, and operation of the 
Bionano Genomics® Saphyr® System (PN 60325). Read this document before performing any procedures on the 
system. 

If the equipment is used in a manner not specified by the manufacturer, the protection provided by the equipment 
may be impaired. 

Important 

You must read the Saphyr Site Preparation Guide and Saphyr System User Guide before plugging the instrument 
into the power supply. 

Before the installation day, we recommend that you familiarize yourself with the relevant safety and installation 
instructions provided in the following documents. 

These documents contain information and warnings that must be followed by the customer for safe operation and 
for keeping the product in a safe condition. Before use of the instrument, please review all safety instructions to 
avoid injury and prevent damage to this instrument and any products connected to it. 

The following documents are available for download fro

www.bionanogenomics.com

. 

Resource 

Description 

Saphyr Site Preparation Guide 

Provides specifications for laboratory space, 
electrical requirements, and environmental 
considerations. 

Saphyr System User Guide 

Provides an overview of instrument components and 
software, proper maintenance, and troubleshooting. 

Bionano Access Software Guide 

Provides an overview of data analysis. 

Safety Warnings 

Make sure that all personnel are trained in the correct operation of the instrument and any potential safety 
considerations. 

Type 

Description 

 

Hazard: Indicates a potential electrical shock hazard 
that may lead to personal injury 

 

Hazard: Indicates a potential laser exposure hazard 
that may lead to personal injury 

Summary of Contents for Saphyr 60325

Page 1: ...r Research Use Only Not for use in diagnostic procedures Copyright 2019 Bionano Genomics Inc All Rights Reserved Saphyr System Safety Guide for Saphyr P N 60325 Document Number 30253 Document Revision...

Page 2: ...ety Guide 5 Important 5 Safety Warnings 5 Laser Safety Warnings 6 Preventing Radiation Exposure 6 Safety Labeling 6 Sound Level Safety Warnings 7 LED Exposure Warning 7 Electrical Safety Warnings 7 El...

Page 3: ...RESPECT TO THIS DOCUMENT EXPRESSED OR IMPLIED INCLUDING BUT NOT LIMITED TO THOSE OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE TO THE FULLEST EXTENT ALLOWED BY LAW IN NO EVENT SHALL BIONANO G...

Page 4: ...30253 Rev A Saphyr System Safety Guide for Saphyr P N 60325 Page 4 of 9 Revision History Revision Notes A Initial Release...

Page 5: ...ormation and warnings that must be followed by the customer for safe operation and for keeping the product in a safe condition Before use of the instrument please review all safety instructions to avo...

Page 6: ...uct is hazardous if the eye is directly exposed to the laser However the diffuse reflections are not harmful Protective eye wear is required when a person is viewing a class 3B laser beam All Class 4...

Page 7: ...ards Only Bionano Field Service Engineers or Bionano certified personnel are qualified to replace the internal fuses The power entry module includes two input fuses on the high voltage input lines Und...

Page 8: ...103 kg 227 lb Environmental Considerations This instrument is designed for indoor use only Element Specification Temperature Maintain a lab temperature of 19 C 66 F to 25 C 77 F Humidity Maintain a n...

Page 9: ...requently asked questions and other related documents from the Support website or by request through e mail and telephone Type Contact Email support bionanogenomics com Phone Hours of Operation Monday...