background image

 

 

30253 Rev A, Saphyr System Safety Guide for Saphyr P/N 60325 

Page 6 of 9 

Laser Safety Warnings 

The Saphyr system has a laser system classification of a Class 1 product, which uses embedded Class 3B and 
one Class 4 laser. The Saphyr System is certified to the IEC 60825-1:2014 standards. 

The nominal wavelength outputs include the following optical power: 

 

500 mW (532 nm) 

 

140 mW (637 nm, 2 each) 

 

400 mW (488 nm) 

 

90 mW (785 nm) 

A Class 1 laser product is safe under all conditions of normal use. 

A Class 4 laser product is hazardous if the eye is directly exposed to the laser. However, the diffuse reflections 
are not harmful. Protective eye wear is required when a person is viewing a class 3B laser beam. All Class 4 
lasers must be equipped with a key switch and a safety interlock. 

Preventing Radiation Exposure 

 

Do not open the front panel of the instrument because you could come into contact with a Class 4 laser. 

 

If you operate the instrument with any panels removed, you risk radiation exposure from the direct or 
reflected laser light. 

 

If you use controls or make adjustments that are not specified in The Saphyr Safety Guide, you can be 
exposed to hazardous radiation. 

Safety Labeling 

Figure 1: Class 4 Laser Warning 

 

 

 

Summary of Contents for Saphyr 60325

Page 1: ...r Research Use Only Not for use in diagnostic procedures Copyright 2019 Bionano Genomics Inc All Rights Reserved Saphyr System Safety Guide for Saphyr P N 60325 Document Number 30253 Document Revision...

Page 2: ...ety Guide 5 Important 5 Safety Warnings 5 Laser Safety Warnings 6 Preventing Radiation Exposure 6 Safety Labeling 6 Sound Level Safety Warnings 7 LED Exposure Warning 7 Electrical Safety Warnings 7 El...

Page 3: ...RESPECT TO THIS DOCUMENT EXPRESSED OR IMPLIED INCLUDING BUT NOT LIMITED TO THOSE OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE TO THE FULLEST EXTENT ALLOWED BY LAW IN NO EVENT SHALL BIONANO G...

Page 4: ...30253 Rev A Saphyr System Safety Guide for Saphyr P N 60325 Page 4 of 9 Revision History Revision Notes A Initial Release...

Page 5: ...ormation and warnings that must be followed by the customer for safe operation and for keeping the product in a safe condition Before use of the instrument please review all safety instructions to avo...

Page 6: ...uct is hazardous if the eye is directly exposed to the laser However the diffuse reflections are not harmful Protective eye wear is required when a person is viewing a class 3B laser beam All Class 4...

Page 7: ...ards Only Bionano Field Service Engineers or Bionano certified personnel are qualified to replace the internal fuses The power entry module includes two input fuses on the high voltage input lines Und...

Page 8: ...103 kg 227 lb Environmental Considerations This instrument is designed for indoor use only Element Specification Temperature Maintain a lab temperature of 19 C 66 F to 25 C 77 F Humidity Maintain a n...

Page 9: ...requently asked questions and other related documents from the Support website or by request through e mail and telephone Type Contact Email support bionanogenomics com Phone Hours of Operation Monday...

Reviews: