30253 Rev A, Saphyr System Safety Guide for Saphyr P/N 60325
Page 8 of 9
Ensure the instrument is installed on a clean, sturdy, level, bench without exposure to direct sunlight or heat
sources.
Do not set the instrument up where liquid or chemicals are used. Ensure that no liquids are spilled into the
instrument.
Do not operate the instrument in the presence of flammable gases or fumes.
Dimensions
Measurement
Instrument
Height
38 cm (15 in)
Width
86 cm (34 in)
Depth
71 cm (28 in)
Weight
103 kg (227 lb)
Environmental Considerations
This instrument is designed for indoor use only.
Element
Specification
Temperature
Maintain a lab temperature of 19°C (66°F) to 25°C
(77°F).
Humidity
Maintain a noncondensing relative humidity between
20–80%.
Elevation
The instrument may only be operated at a location
where the elevation is 2,000m (6,500 ft) or less above
mean sea-level.
Ventilation
Provide at least 5 cm (2 in) of clearance behind the
instrument to allow sufficient ventilation and access
to instrument power connection.
Overhead clearance required for installation and
service is 93 cm (37 inch).
Air Quality
Operate the instrument in a Pollution Degree II
environment or better.