Ensure the instrument is installed on a clean, sturdy, level, bench without exposure to direct sunlight or heat 
sources. 

Do not set the instrument up where liquid or chemicals are used. Ensure that no liquids are spilled into the 
instrument. 

Do not operate the instrument in the presence of flammable gases or fumes. 

Dimensions 

Measurement 

Instrument 

Height 

38 cm (15 in) 

Width 

86 cm (34 in) 

Depth 

71 cm (28 in) 

Weight 

103 kg (227 lb) 

Environmental Considerations 

This instrument is designed for indoor use only. 

Element 

Specification 

Temperature 

Maintain a lab temperature of 19°C (66°F) to 25°C 
(77°F). 

Humidity 

Maintain a noncondensing relative humidity between 
20–80%. 

Elevation 

The instrument may only be operated at a location 
where the elevation is 2,000m (6,500 ft) or less above 
mean sea-level.  

Ventilation 

 

Provide at least 5 cm (2 in) of clearance behind the 
instrument to allow sufficient ventilation and access 
to instrument power connection. 

Overhead clearance required for installation and 
service is 93 cm (37 inch). 

Air Quality 

 

Operate the instrument in a Pollution Degree II 
environment or better. 

 

Summary of Contents for Saphyr 60325

Page 1: ...r Research Use Only Not for use in diagnostic procedures Copyright 2019 Bionano Genomics Inc All Rights Reserved Saphyr System Safety Guide for Saphyr P N 60325 Document Number 30253 Document Revision...

Page 2: ...ety Guide 5 Important 5 Safety Warnings 5 Laser Safety Warnings 6 Preventing Radiation Exposure 6 Safety Labeling 6 Sound Level Safety Warnings 7 LED Exposure Warning 7 Electrical Safety Warnings 7 El...

Page 3: ...RESPECT TO THIS DOCUMENT EXPRESSED OR IMPLIED INCLUDING BUT NOT LIMITED TO THOSE OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE TO THE FULLEST EXTENT ALLOWED BY LAW IN NO EVENT SHALL BIONANO G...

Page 4: ...30253 Rev A Saphyr System Safety Guide for Saphyr P N 60325 Page 4 of 9 Revision History Revision Notes A Initial Release...

Page 5: ...ormation and warnings that must be followed by the customer for safe operation and for keeping the product in a safe condition Before use of the instrument please review all safety instructions to avo...

Page 6: ...uct is hazardous if the eye is directly exposed to the laser However the diffuse reflections are not harmful Protective eye wear is required when a person is viewing a class 3B laser beam All Class 4...

Page 7: ...ards Only Bionano Field Service Engineers or Bionano certified personnel are qualified to replace the internal fuses The power entry module includes two input fuses on the high voltage input lines Und...

Page 8: ...103 kg 227 lb Environmental Considerations This instrument is designed for indoor use only Element Specification Temperature Maintain a lab temperature of 19 C 66 F to 25 C 77 F Humidity Maintain a n...

Page 9: ...requently asked questions and other related documents from the Support website or by request through e mail and telephone Type Contact Email support bionanogenomics com Phone Hours of Operation Monday...

Reviews: