7
1
Overview
The
Waterlase
iPlus tissue cutting system is a
unique device with diverse, hard- and soft-tissue
dental applications. It utilizes advanced laser
and water atomization technologies to safely
and effectively cut, shave, contour, roughen,
etch, and resect oral hard-tissue, and direct laser
energy to perform oral soft-tissue removal,
incision, excision, ablation and coagulation.
Waterlase
iPlus may
also be used for specific
endodontic and periodon
-tal applications.
When used for oral hard-tissue procedures, this
YSGG solid-state laser provides optical energy
to a user-controlled distribution of atomized water
droplets and hydrated surface layer of hard-tissue.
Water present in the target tissue absorbs laser ra-
diation, resulting in explosive molecular expansion
and ablation of hard-tissue. The water in the spray
provides cooling and hydration for the target tissue.
For oral soft-tissue procedures, the
Waterlase
iPlus laser applies optical energy to the soft-tissue
for tissue removal, incision, excision, ablation,
and coagulation using direct laser energy, either
with water for cooling and hydration, or without
water for coagulation.
A flexible Fiber Cable connects at one end to the
laser and at the other to a Handpiece that delivers
laser energy to the target tissue through a Fiber
Tip. A visible light emitted from the Handpiece
head illuminates the area of treatment. The optical
power output and atomized water spray may be
adjusted to specific user requirements for both
soft- and hard-tissue applications.
Waterlase
iPlus is indicated for professional use
on adult and pediatric dental patients. Procedures
must be performed only by licensed dental
practitioners in a dental facility. Use of this device
requires proper clinical and technical proficiency,
and this User Manual provides instructions for use
for those professionals who have completed the
appropriate training.
When used and maintained properly, the
Waterlase
iPlus proves a valuable addition to a practice.
Please contact BIOLASE Customer Service at
1-800-321-6717 in the U.S and Canada for any
service needs; if the user is located outside the
U.S., please contact the BIOLASE-authorized
representative.
This device must be installed, operated, and
maintained according to the guidelines of
CAN/CSA-Z386-14, Safe Use of Lasers in
Healthcare.
Water
l
ase iPlus
User Manual
Summary of Contents for Waterlase iPlus
Page 1: ...User Manual...
Page 2: ...PAGE INTENTIONALLY LEFT BLANK...
Page 62: ...Appendix A Labels Laser Explanatory Label Laser Aperture Label 62 Waterlase iPlus User Manual...
Page 78: ...Appendix G Electromagnetic Compatibility 78 Waterlase iPlus User Manual...
Page 79: ...79 Appendix G Electromagnetic Compatibility Waterlase iPlus User Manual...
Page 80: ...Appendix G Electromagnetic Compatibility 80 Waterlase iPlus User Manual...
Page 81: ...PAGE INTENTIONALLY LEFT BLANK...