Operation Manual for Oscillating Dry Bath                                                          Chapter IV Operation Guide

 

 

8

 — 

In  the  section  information  editing  interface, 

click  on  the 

  option  box  behind  the 

temperature  to  enter  the  temperature  rise  /  drop 

and  speed  setting  interface.  Set  the  temperature 

rise / drop and speed with the 12 keys on the right. 

Click  "OK"  to  set  the  speed  successfully,  click 

"Cancel" to abandon the setting, and then return to 

the section information editing interface. 

                                                                                                                                                                                                                           

                                                                                Temperature Rise / Drop and Speed Setting Interface     

 

In  the  section  information  editing 

interface,  click  on  the 

  option  box  behind  the 

speed to enter the oscillation setting interface. The 

oscillation  modes  include  Forward,  Reversal, 

Interval,  Alternate

.

  The  oscillation  time  and  pause 

time  are  required  to  be  set  when  the  intermittent 

oscillation  is  selected.  Select  the  corresponding 

time and set the time with the 12 keys on the right. 

Click  "OK"  to  set  successfully,  click  "Cancel"  to 

abandon the  setting,  and  then  return to  the  section 

information editing interface.                                                       

                                                                                            Oscillation Setting Interface   

Forward: Clockwise 

Reversal: Counterclockwise 

Interval: Intermittent clockwise or intermittent counterclockwise 

Alternate:  Intermittent  clockwise  and  counterclockwise,  that  is  clockwise,  intermittent,  then 

counterclockwise, with this cycle. 

 

In  the  file  editing  interface,  click  "Save"  to 

enter the file saving interface. Use the 12 keys on 

the right to set the file name. The file name is up 

to  8  digits.  Click  "Back"  to  return  to  the  file 

editing interface. 


 

                                                                                                                  File Saving Interface   

Click "Save" in the file saving and editing interface, if the file name already exists, then a warning box 

Summary of Contents for ThermoShaker MB-202

Page 1: ...ThermoShaker Mixing Block MB 202 Type Oscillating Dry Bath Operation Manual H a n g z h o u B i o e r T e c h n o l o g y C o L t d...

Page 2: ...l at any time without notice All rights reserved Without prior written consent from Hangzhou Bioer Technology Co Ltd any part of the manual shall not be duplicated reproduced or translated into any ot...

Page 3: ...use a The ground connection To avoid the electric shock the input power line must be reliably grounded The instrument is equipped with the three pin plug that has the third pin the pin connects the gr...

Page 4: ...ect shall not be less than 25cm Also don t use the instrument on loose or soft surface or the air inlet at instrument bottom might be blocked Excessive temperature would impair the test performance or...

Page 5: ...by material and manufacturing defects within 12 months after goods delivery Hangzhou Bioer Technology Co Ltd is responsible for selectively repairing and replacing instruments with confirmed defects w...

Page 6: ...AND STORAGE CONDITIONS 2 3 BASIC PARAMETERS AND PERFORMANCE 2 CHAPTER III PREPARATORY WORK 4 1 STRUCTURE DIAGRAM 4 CHAPTER IV OPERATION GUIDE 5 EXAMINATION BEFORE STARTUP 5 CHAPTER V FAULT ANALYSIS A...

Page 7: ...nd culture of all kinds of molecular biological experiments with oscillation requirements etc This product has the following characteristics LCD true color high brightness display Simultaneous display...

Page 8: ...ient temperature 10 C 30 C Relative humidity 70 Power supply AC220V 22V 1 5A 50Hz 1Hz Surface for placement Horizontal stable Caution Before using the instrument please confirm whether the working con...

Page 9: ...25 to 95 Drop 10 from the room temperature Drop from 95 to 25 A 384 12min 12min 12min 3000rpm 0 36 Yes B 96 0 2mL 12min 12min 12min 2000rpm 0 33 Yes C 54 0 5 mL 12min 12min 12min 2000rpm 0 43 None D 3...

Page 10: ...tructure operating keyboard and function of each button of the instrument as well as the preparatory work before the startup When using this instrument for the first time familiarize yourself with the...

Page 11: ...ss the module button the module will be set up and separated from the host Examination Before Startup Please confirm the following content before plugging the plug into the power socket 1 Whether the...

Page 12: ...will enter the main interface The main interface includes Program Single Point Set System Inching these four option buttons Main Interface 2 Program editing Click the Program option in the main inter...

Page 13: ...than or equal to 2 click Delete to delete the selected section File Editing Interface In the file editing interface click Add to add a section after the selected section and enter the section informat...

Page 14: ...on time and pause time are required to be set when the intermittent oscillation is selected Select the corresponding time and set the time with the 12 keys on the right Click OK to set successfully cl...

Page 15: ...Guide 9 will pop up Select Yes to overwrite the existing file select No to go back to the file saving interface and finally save the file after modifying the file name After the file is saved success...

Page 16: ...interface The running interface displays the information of the running section including which section is currently running the current temperature the remaining constant temperature time of the sec...

Page 17: ...that the program running has finished Click OK in the End of Run prompt box to return to the main interface 5 System settings In the main interface click System to enter the system settings interface...

Page 18: ...s Firmware bin After the above operation a firmware upgrade U disk for the oscillating dry bath is made successfully Insert the firmware upgrade U disk into the USB interface of the instrument press a...

Page 19: ...t the time of oscillating operation the instrument body vibration is very large accompanied by abnormal sound The wrong module parts are applied Replace the module parts Contact the supplier or manufa...

Page 20: ...r manufacturer The fan is damaged or not running 7 The module is neither heated nor cooled The temperature sensor is damaged Contact the supplier or manufacturer The cooler is damaged 8 The vibration...

Page 21: ...15 Appendix 1 MB 202 Wiring Diagram...