![biodex 950-385 Application/Operation Manual Download Page 42](http://html1.mh-extra.com/html/biodex/950-385/950-385_application-operation-manual_2753542042.webp)
NOTE: This MEDICAL ELECTRICAL EQUIPMENT needs special precautions regarding EMC and
needs to be installed and put into service according to the EMC information provided in the
ACCOMPANYING DOCUMENTS.
NOTE: Portable and mobile RF communications equipment can affect MEDICAL
ELECTRICAL EQUIPMENT.
NOTE: Contact Biodex Medical Systems for additional EMC information.
ELECTROMAGNETIC COMPATIBILITY FOR GAIT TRAINER 2
Standard
Test Method
Range
Limits
Result
IEC 61000-3-2
Harmonics
100 Hz to 2KHz
Class A
THD = 57.54%
IEC 61000-3-3
Flicker
observation time (TP) 10 min
-
-
max voltage change (dmax)
4%
.45%
max Rel steady state voltage
3%
.14%
change (dc) duration of d(t)>3%(t)
.2 sec
0.00 sec
short term flicker Sev (PST)
1.00
.24
long term flicker Sev (PLT)
.65
.23
IEC 61000-4-2
Electrostatic
Discharge
Contact: 4 & 6Kv pos/neg 1pps
no degradation
complied
for 10 sec
of performance
Air: 2, 4, & 8Kv pos/neg 1pps
for 10 sec
IEC 61000-4-3
Radiated
Immunity
80 MHz to 2500 MHz / 3v/m Horiz
no degradation
complied
& Vertical @ 1M
of performance
IEC 61000-4-4
Electrical Fast
PWR Input leads .5, 1, & 2
no degradation
complied
Transient/Burst, Kv / pos & neg / 5KHz
of performance
Power Leads
Rep Rate
IEC 61000-4-6
Conducted
150 KHz to 80 MHz /3Vrms
no degradation
complied
Immunity,
of performance
Power Leads
IEC 61000-4-8
Magnetic
3A/M RMS @ 50/60 Hz
no degradation
complied
Immunity
of performance
IEC 61000-4-11
Voltage Dips
and Interrupts
Int
Pause
%
no degradation
complied
Duration
between
reduction of performance
Int 20msec
pause 10 sec >95%
no degradation
complied
of performance
Int 100msec
pause 10 sec
60%
no degradation
complied
of performance
Int 500msec
pause 10 sec
30%
no degradation
complied
of performance
Int 5000msec pause 10 sec >95%
no degradation
complied
of performance
CISPR 11
Conducted
150 KHz - 500 KHz /5 MHz - 30 MHz,
79/73 dbuV QP
complied
Emission
Class A, Group 1
66/60 dbuV AV
CISPR 11
Radiated
30 MHz – 230 MHz /230MHz - 1GHz,
40/47 dbuVm
complied
Emissions
Class A, Group 1
@ 10 m
IEC 61000-4-5
Surge
1 Kv, differential, 2 Kv Commons,
no degradation
complied
Immunity,
1ppm 5 Pos,
of performance
Power Leads
5 Neg for total of 10
14. ELECTROMAGNETIC COMPATABILITY
— 14-1 —
ELECTROMAGNETIC COMPATIBILITY
Summary of Contents for 950-385
Page 41: ...Figure 13 1 Installing the optional Elevation Rockers CONTENTS OPTIONAL 13 2...
Page 49: ...17 ASSEMBLY DRAWINGS AND SCHEMATICS 17 1 ASSEMBLY DRAWINGS AND SCHEMATICS...
Page 50: ...98 131 0413 CONTENTS ASSEMBLY DRAWINGS AND SCHEMATICS 17 2...
Page 51: ...CONTENTS 17 3 ASSEMBLY DRAWINGS AND SCHEMATICS...
Page 52: ...CONTENTS ASSEMBLY DRAWINGS AND SCHEMATICS 17 4...
Page 53: ...CONTENTS 17 5 ASSEMBLY DRAWINGS AND SCHEMATICS...
Page 54: ...CONTENTS ASSEMBLY DRAWINGS AND SCHEMATICS 17 6...