CAUTION
ORANGE LED – TROUBLESHOOTING RECOMMENDATIONS
The Mallya base light indicator (LED) remains
orange
if:
a)
The clear square plastic window of the Mallya base and the dose window of the pen
are not correctly aligned, or
b)
The Mallya base is mounted too high / too low
on the pen,
c)
Or, if the Mallya knob is not pushed on enough.
In this case, it is necessary to:
a)
Check that the Mallya window outline lines
up with the pen dose window EXACTLY:
-The pen pointer should fit into the slot on
the Mallya base.
-The notch in the pen dose window should
line up with the notch on the Mallya window
outline.
b)
Check that the top edge of the Mallya base
lines up with the edge of the pen body, so
that it does not sit below or above the edge.
The Mallya base should not cover any of the
pen dose dial.
c)
Press the Mallya knob onto the pen dose dial
so that it almost touches the Mallya base.
When the positions of the Mallya knob and button are correct:
- the Mallya base LED flashes
green
and
orange
if Mallya is not paired,
- the Mallya base LED is
green
if Mallya is paired.