IFU_MAL_US_V1.13_January 2021
P a g e
32 | 44
8.4 COLLECT YOUR INJECTION DATA FROM MALLYA AFTER SMARTPHONE
DISCONNECTION
Your smartphone may disconnect from the Mallya system if Bluetooth is out of range, or the
battery is empty, or for other potential reasons. In this case, you can still use Mallya to record
your insulin doses and, on the next connection, data will be automatically transferred from
Mallya to your smartphone.
Upon re-connection, 2 options are possible:
- Your Mallya application will display the recorded injections data as usual if no
removal
event
occurred during the disconnection period (eg.: your base and / or the knob have / has
not been removed from the original pen in the meantime), or
- Your Mallya application will display the 2 options indicated in
step 8.3.3.
if a removal event
occurred (the Mallya base and / or the knob have / has been removed from the original pen
during the disconnection period).
Follow the instructions in
step 8.3.3.
to indicate:
the pen model on which your Mallya is currently assembled, and
the pen model used for the injections (or primings) during the disconnection period.
CAUTION
Upon a first injection following a disassembly, make sure you reconnect the device
with the application as soon as possible to avoid
“Insulin undetermined”
doses.
Example with
the pen model
"APIDRA"
selected (used
for the injection)