Page 34 of 42
Lamp replacement
Starter replacement
8.3.2 Starter replacement (P/N 100-1370)
•
Turn off the power.
•
Disconnect the power cord from the Universal Hood II.
•
Remove the four screws located on the left/right sides of the drawer.
•
Remove the cover with the UV filter by sliding it forward.
•
Place it down on its backside on a non-abrasive surface.
Caution: Do not put the UV filter directly on the bench.
•
To remove the starter, rotate it counter-clockwise and pull out.
•
To replace the starter insert it into the holder and rotate clockwise.
•
Reassemble the cover and retighten the screws on both sides.