10. Remove the sample from the transilluminator. Remove any liquid or residue remaining
from the sample platen. Place the Fluorescent Reference Plate on the platen. Make sure
that the plate covers the entire glass surface. Close the door and then click on “Continue”
in the window shown above.
11. The next prompt will direct you to turn ON the UV transilluminator. Press the UV
transilluminator or White transilluminator button on the hood, as the case may be, and
click on “Continue”.