Timer delay
To delay the automatic switching off of the instrument, press
for 4 seconds. When the switching off time will
flash, it will be possible to modify it by pressing
.
This operation does not modify the preset switching off time.
Timer esclusion (manual operating)
To operate the instrument manually, it is necessary to set all days to ON. (see “Setting the timer”). On the display
will appear ON. To start or switching off the instrument it will be necessary to use the general switch every time.
Cleaning
Before effecting the cleaning, switch off the instrument and unplug the cable from the socket. Use only alcohol or
non-aggressive detergents. Do not use abrasive products or acids in order to avoid ruining the varnished or plastic
parts. A plastic spatula can be used to remove residual paraffin. The paraffin collecting drawer, placed on the frontal
side, must to be emptied daily.
The instrument does not need to be sterilized because is not expected the treatment of fresh samples but only of
histological samples fixed and included in paraffin.
Maintenance
Effect the periodic cleaning.
MON
08:35
18:30
T1
56°
MON
08:35
ON
T1
56°