10
8.13
- Lock the pressure chamber (03) in its position by turning its lock knob (08) fully to the left,
clockwise. If locking is difficult, it is because the plate locking ring (11) and/or the pressure chamber (03)
are out of their correct position (Image 24).
Img. 24
Img. 25
8.15 -
Simultaneously with the locking of the pressure chamber (03), the activation of the air valve will
occur (with its characteristic click) and the pressurization of the preheated plate. During the
pressurization time, the characteristic sound of compressed air being released into the pressure
chamber will be heard.
8.16 -
After approximately 15 to 20 seconds, depending on the thickness of the plate used,
turn off the main switch (06) to depressurize the pressure chamber (03). At this moment the air valve will
be closed (with its characteristic click) and the flow of compressed air will be interrupted. After complete
depressurization, turn the pressure chamber lock knob (08) fully to the right, counterclockwise, to
release the pressure chamber opening (03) (Image 25).
Attention:
Never force the pressure chamber lock knob (08) to try to open the pressure chamber
(03) before interrupting the flow of compressed air, by turning off the main switch (06).
Failure to follow this safety procedure could result in accidents causing injury to users
and damage to the product.
Attention:
Only activate the pressure chamber lock knob (08) when the pressure chamber (03) is
correctly positioned in the closed position, thus avoiding the escape of compressed air
directly into the environment.
Img. 23