14.4.6 Messages referring to temperature sensor failure
The alarms occur immediately.
•
Audible alarm (buzzer)
•
Switching the zero-voltage relay alarm output
Failure of the temperature sensor for interior measurement:
•
Alarm message “Inner temp. sensor”, collective alarm icon
•
Cause: defective inner temperature sensor
•
Temperature control continues using the safety controller temperature sensor
Failure of the safety controller temperature sensor:
•
Alarm message “Safety control sensor”, collective alarm icon
•
Cause: defective safety controller temperature sensor
Failure of both the temperature sensors for interior measurement and the safety controller:
•
Temperature display shows „- - - - “ or „<-<-<“ or „>->->“
•
Alarm messages “Inner temp. sensor” and “Safety control sensor” alternating, collective alarm icon
•
Cause: both the temperature sensors for interior measurement and the safety controller are defective
•
Refrigeration is turned on permanently.
Failure of temperature sensors of the refrigerating machine:
•
Alarm message “Condensate temp. sensor”, collective alarm icon
Cause: defective temperature sensor on the condenser of the refrigerating machine
•
Alarm message “Cascade temp. sensor”, collective alarm icon
Cause: defective temperature sensor on the cascade of the refrigerating machine
•
Alarm message “Ambient temp. sensor”, collective alarm icon
Cause: defective temperature sensor for ambient temperature or air intake of the 1
st
stage cooling
•
Refrigeration is turned on permanently.
Actions:
•
Please contact BINDER Service.
•
If the temperature rises, i.e. the refrigerating machine is defective (safety controller temperature alarm
and / or tolerance range alarm):
•
Transfer the material to another freezer.
•
Turn off the freezer.
•
If necessary, clean and disinfect the freezer.
Summary of Contents for ULTRA.Guard UF V 500
Page 106: ...UF V E3 09 2018 Page 106 116 27 7 Dimensions UF V 500 E3 Dimensions in mm...
Page 107: ...UF V E3 09 2018 Page 107 116 27 8 Dimensions UF V 700 E3 Dimensions in mm...
Page 109: ...UF V E3 09 2018 Page 109 116...
Page 110: ...UF V E3 09 2018 Page 110 116 29 Product registration...
Page 117: ......