UF V (E3) 09/2018
Page 101/116
27.4 Technical data
Chamber size
500
700
Exterior dimensions
Width, gross (including hinges, door lock and controller
housing)
mm /
inch
920 /
36.2
1204 /
47.4
Height, gross (incl. castors)
mm /
inch
1966 /
77.4
1966 /
77.4
Depth, gross (including power connection, without door
lock and controller housing (equals depth when door
open))
mm /
inch
850 /
33.5
850 /
33.5
Depth, gross (including power connection, door lock and
controller housing)
mm /
inch
1005 /
39.6
1005 /
39.6
Wall clearance rear (minimum)
mm /
inch
100 /
3.9
100 /
3.9
Wall clearance side (side without hinges) (minimum)
mm /
inch
100 /
3.9
100 /
3.9
Wall clearance side (side with hinges) (minimum)
mm /
inch
245 /
9.6
245 /
9.6
Doors
Number of chamber doors
1
1
Number of compartment doors
2
2
Interior dimensions
Quantity of compartments
2
2
Width of inner chamber
mm /
inch
606
890
Height of interior
mm /
inch
1300 /
51.2
1300 /
51.2
Height of individual compartment (with shelves)
mm /
inch
312-319
12.3-12.6
312-319
12.3-12.6
Depth of inner chamber
mm /
inch
604
604
Interior volume, total
l
491 /
17.3
728 /
25.7
Shelves
Quantity of shelves (regular)
3
3
Quantity of shelves (max.)
13
13
Width of shelf
mm /
inch
580 /
22.8
860 /
33.9
Depth of shelf
mm /
inch
590 /
23.2
590 /
23.2
Permitted max. load per shelf (regular shelf)
kg /
lbs
50 /
110
50 /
110
Permitted max. load of inner chamber bottom
kg /
lbs
50 /
110
50 /
110
Permitted total load
kg /
lbs
200 /
441
200 /
441
Quantity of stainless steel racks per level
4
6
Max. quantity of cryo boxes 50 mm /
2 inch
352
528
Max. quantity of cryo boxes 75 mm /
3 inch
224
336
Temperature data
Control range
°C
-40 up to -86
-40 up to -86
Setting range
°C
20 up to -90
20 up to -90
Temperature uniformity (variation) at -80 °C /
-112 °F
± K
2,5
2,5
Fluctuation
± K
1,5
1,5
Pull-down time from +25 °C /
71.6 °F
to -80 °C /
-112 °F
minutes
360
450
Pull-up time in case of power failure from -80 °C /
-112 °F
to -60 °C /
-76 °F
minutes
230
250
Weight
Weight of the chamber (empty)
kg /
lbs
247 / 545
288 /
635
Summary of Contents for ULTRA.Guard UF V 500