How to calibrate in High Mode
Push the [Set switch]
4 red LEDs
blinks
Marking
Rotate 180
degrees
①
Push the [Set switch] with a stick whose nose is
narrow and round. 4 red LEDs start to blink.
②
Rotate the Level Master for 180 degrees referring to
the markings on base level.
Misalignment and insufficient rotation less than 180
degrees causes the error in calibration process.
Note:
In case of calibration in High Mode, adjust the
levels of base level within 0.08 micron in
advance. You can not conduct the calibration
when levels are over 0.08 micron.