List of spare parts

The following components are available from most
laboratory suppliers. In case of difficulty please 
contact Bibby Scientific Ltd.

Component 

Catalogue No. 

Component   

Catalogue No.

Support stand 

D4000/S 

Reservoir level control 

WLS

Boiler D4000/B 

Boiler 

thermostat 

WT4

Condenser WC48/M2 

Condenser 

thermostat 

W14TS

Heating element 240V 

D240H 

Reservoir level control pressure switch 

DPS

Heating element 220V 

D220H 

On/Off switch 

D4000/SW

Hose kit 

W4000/HK 

Boiler O rings

Top 

WB48/01

Gasket kit 

D4000/GK 

Bottom  

QR38/24

Boiler thermostat tube 

WTT48 

Heater sealing ring 

QR38/24

Condenser thermostat tube 

A515414 

Plastic connecting caps 

QC18/11

Silicone rubber rings 

QR18/7 and QR18/10

Fault finding

In the event of operating difficulties with your Distinction Water Still, 
please consult the following notes. If these fail to identify and remedy 
the fault you are advised to contact your supplier or the 
Technical Service Department of Bibby Scientific Ltd.

Symptom

Cause

Remedy

Heater not working

a) Burnt out heater

Replace heater

b) Failed electricity supply

Restore supply

c) Thermostat operated

Identify and rectify the 
cause and reset thermostat

Water level in boiler

a) Supply of cooling water is

Adjust the supply to

is too low - heater exposed

insufficient

approximately 60 l/hr

b) Boiler drain stopcock left open

Close stopcock

Water level in boiler

a) Flow of cooling water is too high

Adjust the supply to

is too high - e.g. water  surging

approximately 60 l/hr

into condenser

b) Drain is restricted

Ensure drain pipe falls freely
from Still with no kinks

Distillate temperature is

Supply of cooling water is 

Adjust the supply to

too high - e.g. >50°C

insufficient

approx 60 l/hr

Water in boiler is “pumped”

a) Tubing from condenser

Ensure tubing falls freely

out of the boiler to drain

distillate outlet to the reservoir

from Still with no kinks

is constricted

b) Mounting for condenser

Remove obstruction

thermostat is blocked

c) Supply of cooling water

Adjust the supply to 

is insufficient

approximately 60 l/hr

Distillate quality is poor*

Boiler heavily scaled

Clean boiler

Output less than 4 l/hr

Mains electricity supply below

nominal voltage

*Distillate quality when measured by pH or conductivity is greatly 

affected by temperature and the presence of absorbed 
carbon dioxide.

Summary of Contents for Distinction D4000

Page 1: ...Distinction Assembly and operating instructions Instructions de montage et de fonctionnement D4000 D4000 EURO Water Still Distinction...

Page 2: ...ith or be affected by other equipment with similar qualifications We cannot be sure that other equipment used in its vicinity will meet these standards and we cannot guarantee that interference will n...

Page 3: ...n Water Still can be wall or bench mounted Select a convenient location which has access to the following services Electricity Supply Before connection please ensure that the line supply is suitable T...

Page 4: ...struct access to the thermostat reset buttons see figure 1 5 Take the two metal straps and attach the springs to the rear of the cradles on the stand Place the assembled boiler and element onto the cr...

Page 5: ...d use 1 5mm M4 eyelet crimps with a performance conforming to BS 4G178 part 1 1984 and part 2 1986 A suitable crimp tool should be used Operation 1 Turn on the cold water supply and adjust the flow to...

Page 6: ...SCALING OPERATION REMOVE ANY ACID SPILLS IMMEDIATELY Turn on the water supply and fill the boiler to the normal operating level The water will flush the acid into the boiler The water supply should be...

Page 7: ...Failed electricity supply Restore supply c Thermostat operated Identify and rectify the cause and reset thermostat Water level in boiler a Supply of cooling water is Adjust the supply to is too low h...

Page 8: ...de fr quences radio Il ne doit donc pas interf rer avec d autres appareils r pondant aux m mes directives Ne pouvant tre s r que d autres appareils en fonctionment autour du distillateur soint confor...

Page 9: ...2 i Ins rer les boulons au travers du collier m tallique et le placer autour du tube vas du bouilleur ii Enrouler le collier en fibre et l ins rer entre le tube vas et le collier m tallique Le pliage...

Page 10: ...collier plastique fourni 10 Placer l extr mit libre du tuyau vers l vacuation en s assurant qu il ne pr sente aucun coude pouvant perturber l coulement de l eau 11 V rifier que le robinet du c ntroleu...

Page 11: ...pince sertir est n cessaire Mise en route 1 Ouvrir le robinet d alimentation en eau froide et r gler le d bit 1 l mn environ L eau doit circuler travers le r frig rant Attendre que le niveau d eau dan...

Page 12: ...TECTION DES GANTS ET DES MASQUES FACIAUX DEVRONT ETRE UTILISES LORS DES OPERATIONS DE NETTOYAGE NETTOYER IMMEDIATEMENT TOUTE ECLABOUSSURE D ACIDE Ouvrir l alimentation d eau pour remplir le bouilleur...

Page 13: ...tion du distillat vers V rifier que le tuyau de sortie du vers l vacuation le r servoir se fait mal distillat ne pr sente aucun coude b Event obstru par la sonde du D gager l vent thermostat r frig ra...

Page 14: ......

Page 15: ...Notes...

Page 16: ...f bibby scientific fr www bibby scientific com Bibby Scientific Italia Srl Via Alcide de Gasperi 56 20077 Riozzo di Cerro al Lambro Milano Italia Tel 39 0 2 98230679 Fax 39 0 2 98230211 e mail marketi...

Reviews: