3. Installation
3.1 Battery
Handpiece has built-in lithium battery. Battery charges wirelessly through the base.
•
Plug in power adapter and connect it to the charging base
•
Seat handpiece on the charging base
• Blue light will be flashing while device is charging
•
Blue light will stay on when fully charged
3.2 Contra angle
•
The contra angle adopts precision gear transmission, and the transmission ratio
is 6:1.
• Clean and disinfect with disinfectant of neutral PH value before the first use.
•
Clean and disinfect after each patient use
•
Sterilize under 134°C, 2.0~2.3bar (0.20~0.23MPa).
• Contra angle in package only fits ApexPilot G2. It does not fit other handpieces.
Installation of contra angle
•
Align any locating pin of the contra
angle with the positioning slot on the
motor handpiece and push in the
contra angle horizontally.
•
The three locating pins on the
contra angle are inserted into the
three positioning slots on the motor
handpiece.
•
A “click” sound indicates that the
installation is in place.
•
The contra angle can be rotated 360°
freely.
Removal of contra angle
Pull out the contra angle horizontally when the motor handpiece is turned off.
Warning
After installation, give the contra angle a light tug to make sure that it is securely
attached before starting motor.
Stop the motor before plugging in or pulling out the contra angle.