14
5.3 File setting screen
5.4 Operating Screen
Standby screen
File list
Default file size
When motor is running,
screen will show the set
speed and set torque. The
moving bars indicate the
actual torque on the file
during operation.
List of available file for
selection
List of file size for
selected file.
Parameters are pre-set
6. Automatic Reverse Function
Automatic reverse is only available in CW mode.
During operation, the motor is in clockwise rotation.
If the load at file exceeds the preset torque, motor will automatically reverse to
counterclockwise.
Motor will resume clockwise rotation when torque is below preset torque.
Load < preset torque
Clockwise
Load > preset torque
Counterclockwise
Load < preset torque
Clockwise resume
Warning:
In low battery, motor cannot execute automatic reverse. It is advised that the
handpiece should always be charged.
If automatic reverse is triggered frequently in the same operation, motor will overload
and stop. If this occurs, please turn off device and allow motor to cool down.