Technical Description Concerning Electromagnetic Immunity 

Table 2: Guidance & Declaration - electromagnetic immunity

Guidance & Declaration — electromagnetic immunity

The model ApexPilot G2 is intended for use in the electromagnetic environment 

specified below. The customer or the user of the model 

ApexPilot G2 should 

assure that It is used in such an environment.
Immunity test

IEC 60601

test level

Compliance 

level

Electromagnetic

environment - guidance

Electrostatic 

discharge 

(ESD) lEC 

61000-4-2

±8kV contact

±2, ±4, ±8,

±15kV air

±8kV contact

±2, ±4, ±8, 

±15kV

air

Floors should be wood, 

concrete or ceramic tile. 

If floors are covered with 

synthetic material, the 

relative humidity should

be at least 30 %.

Electrical fast 

transient/

burst IEC 

61000-4-4

±2kV for power 

supply lines

±1kV for Input/

output lines

±2kV for power 

supply lines

Voltage should be that 

of a typical commercial 

or hospital environment.

Surge

lEC 61000-

4-5

±0.5, ±1kV line to 

line

±0.5, ±1, ±2kV

line to earth

±0.5, ±1kV line 

to line

±0.5, ±1, ±2kV

line to earth

Voltage should be that 

of a typical commercial 

or hospital environment.

Voltage  

dips, short 

interruptions 

and voltage 

variations on

power supply 

input lines

IEC 61000-

4-11

<5 % UT

(>95% dip in UT.)

for 0.5 cycle

<5 % UT

(>95% dip in UT.)

for 1 cycle

70% UT

(30% dip in UT) for 

25 cycles

<5% UT

(>95 % dip in UT)

for 250 cycles

<5 % UT

(>95% dip in 

UT.) for 0.5 

cycle

<5 % UT

(>95% dip in 

UT.) for 1 cycle

70% UT

(30% dip in UT) 

for 25 cycles

<5% UT

(>95 % dip in 

UT) for 250 

cycles

Voltage should be that 

of a typical commercial 

or hospital environment. 

If the user of the models 

ApexPilot G2 requires 

continued operation 

during power mains 

interruptions, it is 

recommended that the 

models ApexPilot G2 

be powered from an 

uninterruptible power 

supply or a battery.

Power 

frequency 

(50/60 Hz)

magnetic field

lEC 61000-

4-8

30A/m

30A/m

Power frequency 

magnetic fields 

should be at levels 

characteristic of a typical 

location in

a typical commercial or

hospital environment.

NOTE 

UT is the a.c. main voltage prior to application of the test level.

Summary of Contents for ApexPilot G2

Page 1: ...Cordless Endo Motor Instruction for Use...

Page 2: ...ofessionals ApexPilot G2 is mainly used for Endodontic treatments It can be used as an endo motor for preparation and enlargement of root canals 1 3 Contraindication Patient or doctor with a pacemaker...

Page 3: ...3 1 4 Package includes Motor handpiece Contra angle Charging base Nozzle Power adapter O ring Silicon cover Mini 8P cable Instruction manual Diagram of components and control buttons...

Page 4: ...tions causing harm to the patient Use original power adapter Other power adapter will result in damage to lithium battery and control circuit The motor handpiece cannot be autoclaved Use disinfectant...

Page 5: ...20 0 23MPa Contra angle in package only fits ApexPilot G2 It does not fit other handpieces Installation of contra angle Align any locating pin of the contra angle with the positioning slot on the mot...

Page 6: ...the ISO standard ISO standard 2 334 2 350 mm Warning Insert and remove file without holding the push button may damage the chuck of contra angle Removal of file 1 Press and hold the push button 2 Pul...

Page 7: ...ers should follow the file requirement for proper operation Press Main to start motor Operating screen Motor runs with set speed and torque Moving bars from bottom up shows the file torque Press Main...

Page 8: ...ber reaching highest goes back to lowest Vice versa Press P to accept and go to the next available parameter Or press Main to accept and exit without going to next parameter Parameter 2 Blinking is th...

Page 9: ...set parameters Or press Main to accept and exit without selecting parameters Pre set parameters Press to scroll and select file size Press P to accept and back to File lists Or press Main to accept a...

Page 10: ...to accept and exit Dominant Hand Press to switch between Left and Right Press P to accept and go to next setting or Press Main to accept and exit Calibration Note Calibration should be performed when...

Page 11: ...tory setting and setting will then switch back to OFF Restore Defaults OFF Location M0 9 battery level operating mode displayed setting displayed setting Location with file name and size 5 Screen disp...

Page 12: ...alcium hydroxide and other solutions Reciprocating path rotary Reciprocating Displayed setting Speed rpm 100 1200 step50 Torque Ncm 0 4 0 6 0 8 1 0 1 2 1 5 1 8 2 0 2 2 2 4 2 5 2 8 3 0 3 5 4 0 4 5 5 0...

Page 13: ...efault 90 F 20 400 step10 Note For injecting calcium hydroxide and other medicant Recommend the difference between F angle and R angle at least 120 if F angle R angle effective cutting angle on Forwar...

Page 14: ...ring operation the motor is in clockwise rotation If the load at file exceeds the preset torque motor will automatically reverse to counterclockwise Motor will resume clockwise rotation when torque is...

Page 15: ...h the file in the contra angle Accurator A2 alarm goes off ApexPilot G2 screen shows AP and bars filling from top to bottom 7 Apex Locator ApexPilot G2 can be connected to Beyes Accurator A2 apex loca...

Page 16: ...ot G2 motor will respond differently according to the setting Apical Action OFF motor stays same operating mode Reverse motor turns CCW Stop motor stops Auto Start Motor initially stops When apex alar...

Page 17: ...Auto Stop Apical Slow Dwn Motor Status Initial Apex alarm on Apex alarm off CW OFF CW CW CW CW low speed CW Stop CW CW CW low speed CW CW CW stop in 1 sec CW low speed CW stop in 1 sec Reverse CW CCW...

Page 18: ...CCW double beeping CCW CCW double beeping low speed CCW REC OFF REC REC REC Stop REC REC REC REC stop in 1 sec Reverse REC CCW double beeping REC Stop CCW double beeping REC CCW double beeping REC sto...

Page 19: ...TR ATR Stop ATR ATR ATR ATR stop in 1 sec Reverse ATR CCW double beeping ATR Stop CCW double beeping ATR CCW double beeping ATR stop in 1 sec Stop ATR Stop ATR Stop Stop ATR Stop ATR stop in 1 sec on...

Page 20: ...rotating file is stuck at the root canal Incorrect specification setting The load torque of file is too high Choose CCW Mode start the motor handpiece and take the file out 9 Cleaning Disinfection an...

Page 21: ...n Operation Operating Mode Warning 1 Preparation Remove the contra angle from handpiece and base 2 Automated Cleaning with washer disinfector Put the contra angle into the washer disinfector Ao value...

Page 22: ...Replace with new battery 6 Put cover back and screw in Note It is recommended for you to contact your local distributor or manufacturer to replace the battery 5 Sterilization Steam sterilization at 13...

Page 23: ...ess oil from the end part of contra angle using air or keep contra angle in a vertical position for gravity draining Warning Motor handpiece should not be filled with oil Caution Hold the contra angle...

Page 24: ...ty extends only to the first retail purchaser of a product and is not transferable or assignable Replacement components or products may be used and or refurbished components or products provided they...

Page 25: ...use the product 14 4 EXCLUSIVE REMEDY CONSEQUENTIAL DAMAGES DISCLAIMER Beyes obligation under this limited warranty is the repair or replacement of defective parts Beyes shall not be liable for and he...

Page 26: ...240V 50Hz 60Hz Output DC5V 1A Battery capacity 3 7V 2000mAh Wireless charging Freq 112 205kHz max RF output power 9 46dBuA m 3m Torque 0 4 5 Ncm Speed 100 1200 rpm Environmental parameters Working Sto...

Page 27: ...below The customer or the user of the model ApexPilot G2 should assure that it is used in such an environment Emissions test Compliance Electromagnetic environment guidance RF emissions CISPR 11 Grou...

Page 28: ...lEC 61000 4 5 0 5 1kV line to line 0 5 1 2kV line to earth 0 5 1kV line to line 0 5 1 2kV line to earth Voltage should be that of a typical commercial or hospital environment Voltage dips short inter...

Page 29: ...xPilot G2 including cables than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter Recommended separation distance d 1 2 P1 2 d 2 P1 2 d 1...

Page 30: ...ntended for use in electromagnetic environment in which radiated RF disturbances is controlled The customer or the user of the model ApexPilot G2 can help prevent electromagnetic interference by maint...

Page 31: ......

Page 32: ...aplein 10 Ie Verd 2595AA The Hague Netherlands Tel 31645171879 English 31626669008 Dutch Email info beyes ca Website www beyes ca Printed in Canada ENI031 Rev 1 23 03 21 Federal law restricts this dev...

Reviews: