6.2.7 The user is responsible for the sterility of the product for the first cycle and
each further usage as well as for the usage of damaged or dirty instruments where
applicable after sterility.
6.2.8 The water quality has to be convenient to the local regulations especially for
the last rinsing step or with a washer-disinfector.
6.2.9 Do not sterilize the motor handpiece, the AC adapter or the base. After each
use, all the objects that were in contact with infectious agents should be cleaned
using towels impregnated with a disinfecting and detergent solution (a bactericidal,
fungicidal and aldehyde free solution) approved by VAH/DGHM-listing, CE marking,
FDA and Health Canada.
6.2.10 To sterilize the endodontic files, refer to the manufacturer’s instructions for
use.
6.2.11 The contra angle needs to be lubricated after cleaning and disinfection, but
before sterilization.
#
Operation
Operating Mode
Warning
1
Preparation
Remove the
contra angle
from handpiece
and base.
2
Automated
Cleaning
with washer-
disinfector
Put the contra
angle into
the washer
disinfector (Ao
value >3000 or,
at least 5 min at
90°C/194°F)
- Avoid any contact between the
contra angle and any instruments, kits,
supports or container.
- Follow instructions and observe
concentrations given by the
manufacturer (see also general
recommendations).- Use only approved
washer-disinfector according to EN
ISO 15883, maintain and calibrate it
regularly.
- Make sure the contra angle is dry
before moving to the next step.
3
Inspection
Inspect the
contra angle and
sort out those
with defects.
- Dirty contra angle must be cleaned and
disinfected again.
- Lubricate the contra angle with an
adequate spray before packaging.
4
Packaging
Pack the
contra angle in
"Sterilization
pouches".
- Check the validity period of the pouch
given by the manufacturer to determine
the shelf life.
- Use packaging which is resistant to a
temperature up to 141°C (286°F) and in
accordance with EN ISO 11607.
9.3 Step-by-Step Procedure