22
Table 3: Guidance & Declaration - electromagnetic immunity concerning
Conducted RF & Radiated RF
Guidance & Declaration - Electromagnetic immunity
The model ApexPilot G1 is intended for use in.the electromagnetic environment
specified below. The customer or the user of the models ApexPilot G1 should
assure that it is used in such an environment.
Immunity test
IEC 60601
test level
Compliance level Electromagnetic environment
- guidance
Conducted RF
lEC 61000-4-6
Conducted RF
lEC 61000-4-6
Radiated RF
lEC 61000-4-3
3 Vrms
150 kHz to
80 MHz
6 Vrms ISM
frequency
band
3 V/m
80 MHz to
2.7 GHz
3V
6V
3V/m
Portable and mobile RF
communications equipment
should be used no closer to
any part of the models Endo
Smart, including cables, than
the recommended separation
distance calculated from the
equation applicable to the
frequency of the transmitter.
Recommended separation
distance
d=1.2×P1/2 d=2×P1/2
d=1.2×P1/2 80 MHz to 800
MHz
d=2.3×P1/2 800 MHz to 2.7
GHz
where P is the maximum
output power rating of the
transmitter In watts (W)
according to the transmitter
manufacturer and d Is the
recommended separation
distance in meters (m).
Field strengths from fixed RF
transmitters, as determined
by an electromagnetic site
survey,a should be less
than the compliance level
in each frequency range.b
Interference may occur In the
vicinity of equipment marked
with the following symbol:
NOTE I
At 80 MHz end 800 MHz. the higher frequency range applies.
NOTE 2
These guidelines may not apply in all situations. Electromagnetic
propagation is affected by absorption and reflection from structures, objects and
people.